Skip to main content
. 2014 Jan;110(1):199–212. doi: 10.1016/j.radonc.2013.11.002

Table 5.

Example 5 – HDR 192Ir source for temporary prostate BT.

Category Typical level (%) Assumptions
Source strength 2 PSDL traceable calibrations
Treatment planning 3 Reference data with the appropriate bin width
Medium dosimetric corrections 1 Full scatter conditions in the pelvic region and for the prostate location are assumed
US-based Treatment planning and delivery: Catheter reconstruction and source positioning accuracy 2 Assuming usage of dedicated catheter reconstruction tools (catheter free-length measurement based methods) for an accurate (0.7 mm) reconstruction of catheter tip and 1.0 mm source positioning accuracy by the afterloader for straight catheters and transfer tubes
US-based 2D and 3D-imaging overall effect 2 US QA performed according to AAPM TG-128 report
Changes of catheter geometry relative to anatomy between intraoperative treatment planning and intraoperative treatment delivery 2 Assuming that new image acquisition and treatment plan calculation is done always before each fraction. It is also required that no manipulation of the implant and anatomy occurs, as it is the case when removing/manipulating the US-probe or moving the patient from the operation table before treatment delivery
Target contouring uncertainty 2 Using CT or CT + T2 imaging
Total dosimetric uncertainty (= 1) 5 For treatment delivery without patient movement and changes in the lithotomic set-up and with the US probe at the position of the acquisition (transversal plane at the prostate base)