Skip to main content
. Author manuscript; available in PMC: 2014 Mar 31.
Published in final edited form as: Blood Press Monit. 2007 Feb;12(1):1–8. doi: 10.1097/MBP.0b013e3280858d5b

Table 2.

Colin 8800 validation study sample compared with AAMI requirements

AAMI SP10: 2002 standards Colin 8800C validation study

Sample size n ≥ 85 (255 data points) n = 96 (288 data points)
Blood pressure
 < 100 mmHg systolic At least 10% of sample 10%
 > 160 mmHg systolic At least 10% of sample 18%
 < 60 mmHg diastolic At least 10% of sample 10%
 > 100 mmHg diastolic At least 10% of sample 10%
 Mean SBP (SD) 134 (28)
 Mean DBP (SD) 76 (15)
 Range for SBP 89–222
 Range for DBP 50–120
Arm sizes
 < 25 cm At least 10% of sample 11%
 > 35 cm At least 10% of sample 16%
Observer agreement
 Within ± 10 mmHg 100% of sample 100%
 Within ± 5 mmHg At least 90% of sample 98%
Mean (SD) of differences between 5 mmHg ± 8 mmHga SBP 3.9 (6.7)
Electronic device and average of the observers DBP – 2.1 (6.1)
Percentage of differencesb ≤ 5 ≤ 10 ≤ 15
 SBP 58 85 95
 DBP 71 91 96

AAMI, Association for the Advancement of Medical Instrumentation; BHS, British Hypertension Society; DBP, diastolic blood pressure; SBP, systolic blood pressure.

a

Required to rate as a ‘pass’ by AAMI standards.

b

For an ‘A’ rating for the BHS standards, 60% of differences between electronic device and manual must fall within 5 mmHg, 85% within 10 mmHg, and 95% within 15 mmHg; for a ‘B’ rating, 50% of differences between the electronic device and the manual must fall within 5 mmHg, 75% within 10 mmHg, and 90% within 15 mmHg.