. 2014 Mar 26;41(4):042102. doi: 10.1118/1.4868693

Table 4.

Highest sample size required for three- or one-region approach for cardiac volume protocol performed on 320-row scanner, as a function of anticipated effective dose, precision, and desired confidence level. Sample sizes displayed are for confidence level of 95%, with sample sizes for 90% confidence level in parentheses.

Precision Anticipated effective dose (mSv)
(q [%]) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
2 8803(7273) 2201(1819) 979(809) 551(455) 353(291) 245(203) 180(149) 138(114) 109(90) 89(73) 73(61) 62(51) 53(44) 45(38) 40(33)
5 498(274) 125(69) 56(31) 32(18) 20(11) 14(8) 11(6) 8(5) 7(4) 5(3) 5(3) 4(2) 3(2) 3(2) 3(2)
10 76(51) 19(13) 9(6) 5(4) 4(3) 3(2) 2(2) 2(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)
15 32(22) 8(6) 4(3) 2(2) 2(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)
20 18(12) 5(3) 2(2) 2(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)
50 3(2) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)