Skip to main content
. 2014 Mar 26;41(4):042102. doi: 10.1118/1.4868693

Table 5.

Highest sample size required for three- or one-region approach for helical chest protocol performed on 320-row scanner, as a function of anticipated effective dose, precision, and desired confidence level. Sample sizes displayed are for confidence level of 95%, with sample sizes for 90% confidence level in parentheses.

Precision Anticipated effective dose (mSv)
(q [%]) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
2 8402(6941) 2101(1736) 934(772) 526(434) 337(278) 234(193) 172(142) 132(109) 104(86) 85(70) 70(58) 59(49) 50(42) 43(36) 38(31)
5 476(262) 119(66) 53(30) 30(17) 20(11) 14(8) 10(6) 8(5) 6(4) 5(3) 4(3) 4(2) 3(2) 3(2) 3(2)
10 73(49) 19(13) 9(6) 5(4) 3(2) 3(2) 2(1) 2(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)
15 30(21) 8(6) 4(3) 2(2) 2(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)
20 17(12) 5(3) 2(2) 2(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)
50 3(2) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1) 1(1)