Skip to main content
Journal of Optometry logoLink to Journal of Optometry
. 2010 Nov 4;1(2):59–64. doi: 10.3921/joptom.2008.59

Accuracy of Schiotz Tonometry in Measuring the Intraocular Pressure After Corneal Refractive Surgery

Nicola Rosa 1,2,3,, Michele Lanza 1,2,3, Gilda Cennamo 2, Luigi Capasso 1, Gennarfrancesco Iaccarino 1, Maria Borrelli 3, Fabrizia Fusco 1, Antimo Flagiello 1
PMCID: PMC3972649

Abstract

Purpose

To test the accuracy of Schiotz tonometry (ST) in patients that underwent photorefractive keratectomy (PRK).

Methods

175 eyes previously treated with PRK for myopia, myopic astigmatism or mixed astigmatism, with a mean preoperative spherical equivalent (SE) of −4.27±2.83 diopters (range: -13 to −0.25 diopters), had their intraocular pressure (IOP) measured, at least 3 months after PRK, with Goldmann applanation tonometry (GAT) and ST with both 5.5-g and 10-g load and also utilizing the Friedenwald nomogram. These measurements were retrospectively compared to the IOP readings obtained with GAT before surgery.

Results

Before PRK the mean IOP, as measured with GAT, was 14.7±2.63 mmHg. After PRK, the mean IOP readings were 10.9±2.85 mmHg (P<0.001) with GAT, 13.58±3.42 mmHg (P<0.001) with 5.5-g load ST; 12.32±3.67 mmHg (P<0.001) with 10-g load ST, and 16.19±5.63 mmHg (P<0.001) with the Friedenwald nomogram. The different methods showed a poor agreement, the worst one being between GAT before surgery and the Friedenwald nomogram (r2=0.05). Moreover, ST with both 5.5 g and 10 g load showed an underestimation (more pronounced for the 10-g than for the 5.5-g load) that increased with the amount of treatment.

Conclusions

Although ST with a 5.5 g load seems to be more accurate to measure the true IOP in patients that have undergone PRK compared to the Goldmann technique, it cannot be considered to be accurate in a sufficient number of cases. IOP measurement after refractive surgery is still a problem, and further studies are needed to overcome it.

Key Words: Schiotz tonometry, Goldmann applanation tonometry, PRK

Introduction

The measurement of intraocular pressure (IOP) after refractive surgery is one of the most challenging problems. Following a refractive surgery procedure, such as myopic photorefractive keratectomy (PRK), and laser in situ keratomileusis (LASIK), both applanation (GAT) and non-contact (NCT) tonometry have been reported to underestimate the IOP, so different methods such as pneumotonometry or dynamic contour tonometry have been proposed to overcome this problem.1-5

In two case reports,6,7 indentation tonometry with a Schiotz tonometer (ST) has been able to detect increases in IOP after LASIK undetected by GAT. However, excluding these case reports, we were unable to find in the literature (we did a computer search using the PubMed database of the National Library of Medicine) any extensive study proving the real efficacy of this method after refractive surgery.

For this reason we decided to test the accuracy of ST in patients that had undergone PRK.

Methods

One hundred and seventy-five eyes of 105 patients (42 males and 63 females) that had been previously treated with PRK for myopia, myopic astigmatism or mixed astigmatism in our Department were included in this clinical study. The research followed the tenets of the Declaration of Helsinki; informed consent was obtained from the subjects after explanation of the nature and possible consequences of the study; the research was approved by the institutional human experimentation committee.

The age of the patients at the time of refractive surgery ranged from 20 to 59 (mean±SD = 32±8.7 years). All treatments were performed under topical anesthesia with oxybuprocaine eye drops. The lids were kept open with a speculum, the epithelium debrided with a mechanical brush and all treatments performed with a 193 nm excimer laser (Nidek EC 5000, Nidek Co. Ltd, Japan) operating in scanning mode. After treatment a smoothing of the surface was performed using a 0.04% hyaluronic acid solution. A bandage contact lens was applied under sterile conditions on the treated eye immediately following surgery, and was left until complete re-epithelialization. During this period the operated eyes received the following medications: Diclofenac sodium 0.1% eye drops twice a day for the first 2 days, nethylmicin preservative-free eye drops until re-epithelialization and preservative-free artificial tears for 1 month. After re-epithelialization, clobetasone eye drops were prescribed to all patients for 1 month in a tapered dose, as follows: one drop four times a day for the first week, one drop three times a day for the second week, one drop twice a day for the third week, and one drop once a day for the last week.

The IOP was evaluated both with GAT and ST. In particular, GAT was performed first and ST followed 10 minutes later both with a 5.5-g and a 10-g load. The measurements were performed by two different observers, the second one not being aware of the results obtained by the first one.

Both 5.5 and 10 g load were used to calculate the IOP with the Friedenwalt nomogram. These postoperative values were compared with the GAT readings obtained before surgery as these values can be considered IOP real values.

To further validate our study, for those 49 patients that had had surgery only in one eye, we subtracted the change in GAT reading obtained in the fellow non-operated eye, from the change in the treated eye; these values were compared to ST readings. The refraction of the eyes in both groups before and after surgery is summarized in table 1. The refraction of the 49 unoperated fellow eyes ranged from -9 D to + 1 D (mean ±SD = -4.59±2.32 D) of sphere, the cylindrical power from −3 D to 0 D (mean ±SD = -0.83±1.1 D) and the spherical equivalent refraction from -10 D to -1 D (mean ± SD = -3.78±2.81 D).

Table 1.

Refraction (diopters) of the eyes included in groups 1 and 2, both before and after surgery

Before Sugery After Surgery
Group 1 Sphere -4.27±2.83
(from -13 to +1.5)
+0.66±0.77
(from −1.5 to +5)
Cylinder -1.17±1.24
(from -6 to +3.5)
+0.2±0.38
(from –0.5 to +2.5)
Spherical Equivalent -4.86±2.79
(from -13 to -0.25)
+ 0.75±0.81
(from–1.75 to +5.25)
Group 2 Sphere -5.96±3.3
(from -13 to +1)
+1.25±1.16
(from –0.75 to +4.5)
Cylinder -1.07±1.19
(from -4.5 to 3.5)
+0.42±0.61
(from –0.5 to +2.5)
Spherical Equivalent -6.42±3.38
(from -13 to -0.25)
+1.4±1.25
(from –0.75 and +4.75)

Analyses were carried out using Microsoft Excel 7 (Microsoft Corp., Seattle, USA) and GraphPad InStat version 3.0 (GraphPad Software, San Diego, USA). The statistical evaluation was performed with a paired two-tailed Student's T test and with Bland Altmann correlation. For this latter evaluation we decided to ascribe some significance only to changes outside the ±2 mmHg interval, as it had been shown that if single tonometry readings are used by the same observer, they must differ by at least 4.5 mmHg before it can be said that the IOP is different; for variations of smaller magnitude there is only a 10% chance that this difference is due to a change alone.8 Moreover, Whitacre and Stein9 recommended not to ascribe any clinical significance to changes in the IOP measured across two different measurement sessions that were smaller than 2-3 mmHg.

Results

The IOP as measured with GAT before PRK and with GAT and ST after PRK, for group-1 patients, is summarized in table 2. The correlation between the IOP measured before and after surgery with GAT and the ST measurements (5.5 g, 10 g, and nomogram to take into account the variations in scleral rigidity across patients) is shown in Figures 1 to 8.

Table 2.

Ranges, means, standard deviations (SD) and P values between GAT before surgery and GAT, ST 5.5 g, ST 10 g and Fiendewald nomogram after PRK in group 1

Group 1




Range (mmHg) Mean (mmHg) SD P
GAT before 8-22 14.7 2.63
GAT after 5-20 10.9 2.85 <0.001
ST 5.5 g 7.1-24.4 13.58 3.42 <0.001
ST 10 g 4.9-23.1 12.32 3.67 <0.001
Friedenwald 4-31 16.19 5.63 <0.001

Figure 1.

Figure 1

Scatter plot that shows the correlation between IOP values before and after surgery obtained with GAT in group 1 (r2=0.1062, P<0.0001).

Figure 2.

Figure 2

Bland-Altman plot comparing GAT before and after surgery for group-1 patients. Solid line represents the average difference between methods; dotted lines represent 95% Confidence Interval limits. Broken line represents the correlation line (r2=0.0077, P=0.2497).

Figure 3.

Figure 3

Scatter plot that shows the correlation between IOP values before surgery obtained with GAT and IOP measurements after surgery performed with ST 5.5 g for group-1 patients (r2=0.1204, P<0.0001).

Figure 4.

Figure 4

Bland-Altman plot between GAT before surgery and ST 5.5 g after surgery for group-1 patients. Solid line represents the average difference between methods; dotted lines represent 95% the Confidence Interval limits. Broken line represents the correlation line (r2=0.0745, P=0.0003).

Figure 5.

Figure 5

Scatter plot that shows the correlation between IOP values before surgery obtained with GAT and IOP measurements after surgery performed with ST 10 g for group-1 patients. (r2=0.0958, P=0.0006).

Figure 6.

Figure 6

Bland-Altman plot comparing GAT before surgery and ST 10 g after surgery for group-1 patients. Solid line represents the average difference between methods; dotted lines represent the 95% Confidence Interval limits. Broken line represents the correlation line (r2=0.1165, P<0.0001).

Figure 7.

Figure 7

Scatter plot that shows the correlation between IOP values before surgery obtained with GAT and IOP measurements after surgery performed with a Friendewald Nomogram for group-1 patients (r2=0.0535, P=0.0026).

Figure 8.

Figure 8

Bland-Altman plot comparing GAT before surgery and ST 10 g after surgery for group-1 patients. Solid line represents the average difference between methods; dotted lines represent the 95% Confidence Interval limits. Broken line represents the correlation line (r2 = 0.4317, P<0.0001).

In Particular, Bland-Altman plots show that the comparison of ST with preoperative GAT readings (Figures 4, 6 and 8) reveals a proportional error, as the differences are negative in case of low IOP measurements and positives in case of higher IOP values, mainly when we used the Friedenwald nomogram.

Similarly, the values ranges, means, standard deviations and P values corresponding to the differences of GAT before PRK and GAT and ST after PRK for the second group are shown in table 3.

Table 3.

Ranges, means, standard deviations (SD) and P values between the differences of GAT before PRK and GAT, ST 5.5 g, ST 10 g and Fiendewald nomogram after PRK in the second group

Difference with GAT before Surgery in Second Group
Non-modified Group 2 Range (mmHg) Mean (mmHg) SD P
GAT after -15 to +3 -3.9 3.38 <0.001
ST 5.5 g -8.2 to +10.4 -0.28 4.09 0.63
ST 10 g -10.5 to +9.3 -2.17 4.27 <0.001
Friedenwald -11 to +22 +2.21 6.43 0.047

Moreover, for Group 2, at the same follow-up interval of the operated eye (before and at least 3 months after surgery), in the non-operated eye, GAT showed a difference that ranged between -6 mmHg and +5 mmHg (mean ±SD = −0.58±2.21 mmHg), with a non-significant difference (P=0.122).

The same values obtained subtracting the change measured with GAT in the untreated eye from the change in the treated eye are shown in table 4.

Table 4.

Ranges, means, standard deviations (SD) and P values between the differences of GAT before PRK and GAT, NCT, ST 5.5 g, ST 10 g and Fiendewald nomogram after PRK in the modified second group

Difference of IOP before and after Surgery in Second Group
Modified Group 2 Range (mmHg) Mean (mmHg) SD P
GAT after -15 to +3 -4.21 3.47 <0.001
ST 5.5 -10.4 to +8.2 +0.53 4.29 <0.001
ST 10 -9.3 to +10.5 +2.41 4.44 0.39
Friedenwald -17 to +6 -2.69 6.08 <0.001

Table 5 shows the number of patients for whom the postoperative IOP fell in the ±1-mmHg interval and in the ±2-mmHg interval, relative to the preoperative GAT values, for the different methods under study.

Table 5.

Number of patients for whom the postoperative IOP fell in the ± 1 mmHg interval and in the ± 2 mmHg interval relative to the preoperative GAT values, for the different groups and for the different techniques under study

Different Group Ranges
Range GAT Schiotz 5.5 g Schiotz 10 g Friedenwald
Group 1 ±1 mmHg 30 (17.1%) 22 (12.6%) 24 (13.7%) 31 (17.7%)
±2 mmHg 51 (29.1%) 81 (46.3%) 79 (45.1%) 56 (32%)
Group 2 (not modified) ±1 mmHg 10 (20.4%) 12 (24.5%) 5 (10.2%) 4 (8.2%)
±2 mmHg 14 (28.6%) 22 (44.9%) 16 (32.7%) 7 (14.3%)
Group 2 (modified) ±1 mmHg 9 (18.4%) 11 (22.4%) 7 (14.3%) 10 (20.4%)
±2 mmHg 12 (24.5%) 18 (36.7%) 15 (30.6%) 15 (30.6%)

Moreover, ST with both 5.5 and 10 g load showed an underestimation (more pronounced for the 10 g measurements than for the 5.5 g load) that increased with the amount of treatment. On the contrary, when we applied the Friedenwald nomogram such relationship disappeared (Figures 9-11).

Figure 9.

Figure 9

Correlation between differences in IOP measurements (ST 5.5 g load - preoperative GAT) and effective treatment. (r2=0.1046).

Figure 10.

Figure 10

Correlation between differences in IOP measurements (ST 10 g load - preoperative GAT) and effective treatment. (r2=0.1899).

Figure 11.

Figure 11

Correlation between differences in IOP measurements (ST with Friedenwald nomogram - preoperative GAT) and effective treatment. (r2=0.0022).

Discussion

Several authors studying the IOP after refractive surgery found the measurements obtained with both GAT and NCT10-16 unreliable, and proposed different ways to overcome this problem, such as correcting factors,10,11,13 or different techniques, such as pneumotonometry, dynamic contour tonometry or peripheral measurements.1-3,5,17

The possible causes of the decrease of the measured postoperative IOP with GAT and NCT are debated. Mardelli et al.18 proposed that the removal of Bowman's membrane and the deposition of newly synthesized collagen material alter the resistance of the cornea to applanation. In the same way, Patel and Aslanides19 suggested that the general softening of the corneal tissue after the natural healing process following PRK might be the cause of the lower postoperative IOP readings.

On the contrary, Rosa et al.10 and Montés-Micó and Charman12 supported the hypothesis that the decrease in thickness and the decrease in curvature might be the reason behind this underestimation.

One of the main problems in this kind of study is to establish which one is the real IOP.

The best approach should be to compare the IOP readings with the direct measurement of the IOP, but this is very difficult to perform in human eyes. Today, even if it has been shown that the accuracy of GAT depends on many factors including corneal thickness and structure, it is considered the standard for measuring IOP in ophthalmology in normal eyes1 and it is used world-wide to validate other measuring methods.20-23

For this reason, we decided to retrospectively compare the ST measurements with the GAT values obtained preoperatively, assuming that there are no changes in the IOP due to the surgical procedure. One criticism that could be raised against this study is that we did not check the IOP before surgery with ST, which is true. However:

  • 1.

    This is a retrospective study and ST is not used any-more to check the IOP in normal eyes as it is a gross method subject to many sources of error.

  • 2.

    The purpose of the study was not to establish whether ST values correlate with the ST measurements obtained before surgery, but to establish whether ST is able to give IOP measurements after PRK that reflect the real IOP, which was postulated to be the GAT measured preoperatively.

  • 3.

    The validity of our choice was confirmed by the fact that for the 49 fellow unoperated eyes, no significant changes were detected relative to the initial IOP values.

The results of the present study confirm that after PRK, GAT significantly underestimates IOP. Regarding ST, we found the IOP measured with this method to be higher than that measured with GAT. The reason for this is not clear to us, but the different diameter and shape of the plunger might explain this difference, more than the different mechanism (indentation vs. applanation).

This could also explain why when utilizing Friedenwald nomogram we found even higher values. In fact, one possible explanation could be that we measure the corneal rigidity in the area where the maximal ablation is performed, where the cornea is thinner compared to the remaining structure. These findings are confirmed by the Bland-Altman plot (Figure 8), which shows a systematic error caused by a much higher range of values obtained with this device and that it is clear the trend to change bias for values of average IOP below and above 13 to 15 mmHg.

In conclusion, even if a prospective study should be performed to further validate our findings, our results showed such a wide spread of data that less than 50% of the eyes fell in the ±2-mmHg interval relative to the presumed real IOP. For this reason, ST cannot be used alone, but perhaps it could be used in combination with other methods in an attempt to calculate the right value. Further studies are required to establish if such a method could be reliably used in patients that have undergone other refractive surgery procedures, such as LASIK, or may only be limited to particular LASIK cases complicated with interstitial keratitis and accumulation of intrastromal fluid.6,7

References

  • 1.Kaufmann C., Bachmann L.M., Thiel M.A. Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci. 2003;44:3790–3794. doi: 10.1167/iovs.02-0946. [DOI] [PubMed] [Google Scholar]
  • 2.Zadok D., Tran D.B., Twa M. Pneumotonometry versus Goldmann tonometry after laser in situ keratomileusis for myopia. J Cataract Refract Surg. 1999;25:1344–1348. doi: 10.1016/s0886-3350(99)00202-3. [DOI] [PubMed] [Google Scholar]
  • 3.Abbasoglu O.E., Bowman R.W., Cavanagh H.D. Reliability of intraocular pressure measurements after myopic excimer photorefractive keratectomy. Ophthalmology. 1998;105:2193–2196. doi: 10.1016/S0161-6420(98)91215-5. [DOI] [PubMed] [Google Scholar]
  • 4.Stahl J., Vold S. Effect of corneal thickness on the accuracy of intraocular pressure measurement in rabbits after excimer laser photoablation. J Cataract Refract Surg. 1999;26:736–743. doi: 10.1016/s0886-3350(99)00463-0. [DOI] [PubMed] [Google Scholar]
  • 5.Duch S., Serra A., Castanera J. Tonometry after laser in situ keratomileusis treatment. J Glaucoma. 2001;10:261–265. doi: 10.1097/00061198-200108000-00003. [DOI] [PubMed] [Google Scholar]
  • 6.Rehany U., Bersudsky V., Rumelt S. Paradoxical hypotony after laser in situ keratomileusis. J Cataract Refract Surg. 2000;26:1823–1827. doi: 10.1016/s0886-3350(00)00763-x. [DOI] [PubMed] [Google Scholar]
  • 7.Fogla R., Rao S.K., Padmanabhan P. Interface fluid after laser in situ keratomileusis. J Cataract Refract Surg. 2001;27:1526–1528. doi: 10.1016/s0886-3350(00)00881-6. [DOI] [PubMed] [Google Scholar]
  • 8.Motolko M.A., Feldman F., Hyde M. Sources of variability in the results of applanation tonometry. Can J Ophthalmol. 1982;17:93–95. [PubMed] [Google Scholar]
  • 9.Whitacre M.M., Stein R. Sources of error with use of Goldmann type tonometers. Surv Ophthalmol. 1993;38:1–30. doi: 10.1016/0039-6257(93)90053-a. [DOI] [PubMed] [Google Scholar]
  • 10.Rosa N., Cennamo G., Breve M.A. Goldmann applanation tonometry after myopic photorefractive keratectomy. Acta Ophthalmol Scand. 1998;76:550–554. doi: 10.1034/j.1600-0420.1998.760508.x. [DOI] [PubMed] [Google Scholar]
  • 11.Cennamo G., Rosa N., La Rana A. Non-contact tonometry in patients that underwent photorefractive keratectomy. Ophthalmologica. 1997;211:341–343. doi: 10.1159/000310825. [DOI] [PubMed] [Google Scholar]
  • 12.Montés-Micó R., Charman W.N. Intraocular pressure after excimer laser myopic refractive surgery. Ophthalmic Physiol Opt. 2001;21:228–235. doi: 10.1046/j.1475-1313.2001.00581.x. [DOI] [PubMed] [Google Scholar]
  • 13.Chatterjee A., Shah S., Bessant D.A. Reduction in intraocular pressure after excimer laser photo-refractive keratectomy. Correlation with pretreatment myopia. Ophthalmology. 1997;104:355–359. doi: 10.1016/s0161-6420(97)30308-x. [DOI] [PubMed] [Google Scholar]
  • 14.Doughty M.J., Zaman M.L. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44:367–408. doi: 10.1016/s0039-6257(00)00110-7. [DOI] [PubMed] [Google Scholar]
  • 15.Gimeno J.A.S., Munoz L.A., Valenzuela L.A. Influence of refraction on tonometric readings after photorefractive keratectomy and laser assisted in situ keratomileusis. Cornea. 2000;19:512–516. doi: 10.1097/00003226-200007000-00022. [DOI] [PubMed] [Google Scholar]
  • 16.Arimoto A., Shimizu K., Shoji N. Underestimation of intraocular pressure in eyes after laser in situ keratomileusis. Jpn J Ophthalmol. 2002;46:645–649. doi: 10.1016/s0021-5155(02)00555-5. [DOI] [PubMed] [Google Scholar]
  • 17.Schipper I., Senn P., Oyo-Szerenyi K. Central and peripheral pressure measurements with the Goldmann tonometer and Tono-Pen after photorefractive keratectomy for myopia. J Cataract Refract Surg. 2000;26:929–933. doi: 10.1016/s0886-3350(99)00461-7. [DOI] [PubMed] [Google Scholar]
  • 18.Mardelli P.G., Piebenga L.W., Whitacre M.M. The effect of excimer laser photorefractive keratectomy on intraocular pressure measurements using the Goldmann applanation tonometer. Ophthalmology. 1997;104:945–948. doi: 10.1016/s0161-6420(97)30202-4. [DOI] [PubMed] [Google Scholar]
  • 19.Patel S., Aslnides I.M. Main causes of reduced intraocular pressure after excimer laser photorefractive keratectomy. J Refract Surg. 1996;12:673–674. doi: 10.3928/1081-597X-19960901-04. [DOI] [PubMed] [Google Scholar]
  • 20.Troost R., Vogel A., Beck S. Clinical comparison of two intraocular pressure measurement methods: Smartlens dynamic observing tonography versus Goldmann. Graefes Arch Clin Exp Ophthalmol. 2001;239:889–892. doi: 10.1007/s00417-001-0376-4. [DOI] [PubMed] [Google Scholar]
  • 21.Marchini G., Babighian S., Specchia L. Evaluation of the new Ocuton S tonometer. Acta Ophthalmol Scand. 2002;80:167–171. doi: 10.1034/j.1600-0420.2002.800209.x. [DOI] [PubMed] [Google Scholar]
  • 22.Jorge J., Diaz-Rey J.A., González-Méijome J.M. Clinical performance of the Reichert AT550: a new non-contact tonometer. Ophthalmic Physiol Opt. 2002;22:560–564. doi: 10.1046/j.1475-1313.2002.00077.x. [DOI] [PubMed] [Google Scholar]
  • 23.Fresco B.B. A new tonometer--the pressure phosphene tonometer: clinical comparison with Goldman tonometry. Ophthalmology. 1998;105:2123–2126. doi: 10.1016/s0161-6420(98)91137-x. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Optometry are provided here courtesy of Elsevier

RESOURCES