Table 4. Back calculated accuracy and precision levels of calibration standards and spiked samples during the validation.

Analyte Calibration standards Recovery ±s (concentration) Spiked samples Recovery ±s (spiked concentration)a
Low Medium High Low Medium High
p-Tyr 122±4 (0.8 μM) 86±4 (12.5 μM) 95±1 (150 μM) 90±5 (110 μM) 111±3 (140 μM) 123±2 (170 μM)
m-Tyr 98±6 (31 nM) 100±5 (125 nM) 98.9±0.6 (188 nM) 86±1.6 (80 nM) 90.0±0.5 (130 nM) 97±3 (200 nM)
o-Tyr 103±9 (31 nM) 95±3 (250 nM) 99±3 (750 nM) 96±4 (200 nM) 100.2±0.5 (475 nM) 100.7±0.8 (725 nM)
Phe 109±4 (0.4 μM) 100±2 (12.5 μM) 99±1 (150 μM) 98±2 (60 μM) 100.6±0.2 (115 μM) 100.8±0.7 (170 μM)
3NO2-Tyr 104±9 (31 nM) 93±7 (250 nM) 98±3 (750 nM) 77±4 (180 nM) 93±5 (455 nM) 91.6±0.3 (725 nM)
3Cl-Tyr 107±6 (16 nM) 89±3 (250 nM) 99±3 (1500 nM) 99.8±1.5 (180 nM) 100.3±0.8 (910 nM) 101.0±0.9 (1600 nM)
2-dG 109±8 (16 nM) 88±5 (125 nM) 9.70 98±2 (750 nM) 99±1 (250 nM) 103±1 (530 nM) 101±2 (725 nM)
8OHdG 106±10 (4 nM) 88±8 (31 nM) 98±3 (188 nM) 67±3 (80 nM) 81±2 (120 nM) 81±6 (190 nM)
a

: concentrations in the spiked samples were calculated after subtracting the mean levels (n = 3) measured in the non-spiked pooled urine samples during each validation batch.