Skip to main content
. 2014 Winter;13(Suppl):43–50.

Table 1.

Intra- and interday precision and accuracy of cetrorelix QC samples using the described HPLC method.

Theoretical concentration Calculated concentration
(mean±S.D., n=3)
Precision
(R.S.D.) (%)
Accuracy
Recovery %
Intraday
125 121.6 ± 0.11 0.11 96.85
250 242.56 ± 0.40 0.17 97.03
500 485.13 ± 1.28 0.26 97.03
Interday
125 121.28 ± 1.90 1.56 97.02
250 243.46 ± 2.97 1.21 97.38
500 488.56 ± 5.57 1.14 98.69