Two-dimensional representations of the
quantitative biases by (A)
peptide-, extended-peptide-, and protein-level calibration approaches
and (B) the two “hybrid” calibration approaches (c.f.,
Figure 1). QC samples were prepared by spiking
blank plasma with pure protein at three levels: 1.6, 10, and 80 μg/mL.
The purities of all standards were accurately measured by the quantitative
amino acid analysis method to eliminate bias arising from possible
inaccurate purity. Five aliquots of each QC sample were individually
prepared and analyzed in replicates by the five calibration approaches.
Each sample was analyzed three times in each of two different days
(day 1 and day 14, N = 6, shown as individual data
points). For every calibration method, the quantitative values were
obtained independently using the two signature peptides (SP) (i.e.,
the GPS and TVA peptides). The two axes represent the quantitative
biases by the two SP. The red box in the center of each panel denotes
the zone of <20% bias, while the golden box signifies the zone
of <10% bias.