Skip to main content
. 2014 Mar 10;86(7):3575–3584. doi: 10.1021/ac5001477

Table 1. Pharmacokinetic Parameters Calculated Based on the Quantitative Results Obtained by Different Calibration Approaches and Signature Peptides.

  AUC (μg h/mL) Cl (mL/h) Vss (mL) T1/2 (h)
protein calibration, GPS 7274 0.083 28.9 260.7
protein calibration, TVA 7113 0.085 28.8 260.8
aprotein calibration, mean of two SP 7194 0.084 28.9 260.9
protein calibration, deviation between the two SP 2% 2% 0% 0%
hybrid no. 2 (SIL-extended-peptide I.S.), GPS 7551 0.080 30.6 298.5
hybrid no. 2 (SIL-extended-peptide I.S.), TVA 6763 0.089 32.3 263.9
hybrid no. 2, mean of two SP 7155 0.084 31.4 281.2
hybrid no. 2, deviation between the two SP 8% 8% 4% 9%
hybrid no. 1 (SIL-peptide I.S.), GPS 7244 0.083 27.9 252.5
hybrid no. 1 (SIL-peptide I.S.), TVA 7125 0.085 28.9 261.1
hybrid no. 1, mean of two SP 7212 0.084 28.3 252.5
hybrid no. 1, deviation between the two SP 1% 1% 2% 2%
extended-peptide calibration, GPS 5774 0.105 38.2 271.1
extended-peptide calibration, TVA 3696 0.163 49.5 239.2
extended-peptide calibration, mean of two SP 4721 0.128 43.2 255.6
extended-peptide calibration, deviation between SP 31% 31% 18% 9%
peptide calibration, GPS 5640 0.107 35.9 252.5
peptide calibration, TVA 2995 0.202 74.7 280.7
peptide calibration, mean of two SP 4293 0.141 49.0 260.4
peptide calibration, deviation between the two SP 43% 43% 50% 7%
a

Parameters were calculated based on the mean of the concentrations obtained by the two SP.