Skip to main content
. Author manuscript; available in PMC: 2014 May 1.
Published in final edited form as: IEEE Trans Biomed Eng. 2013 Jan 9;60(5):1446–1452. doi: 10.1109/TBME.2012.2237030

Fig. 10.

Fig. 10

The effect of inter-electrode distance on the measured 35 kHz resistance, reactance and phase for 2 different conductivities relative to a baseline conductivity of 0.60 S/m. The changes in surface measured impedance values are only minimally affected by inter-electrode distances, regardless of the value of the conductivity.