Skip to main content
. Author manuscript; available in PMC: 2014 May 1.
Published in final edited form as: IEEE Trans Biomed Eng. 2013 Jan 9;60(5):1446–1452. doi: 10.1109/TBME.2012.2237030

Fig. 4.

Fig. 4

The effect of altering the thickness of muscle on the measured impedance parameters using the electrode configuration and spacing shown in Fig. 1. As can be seen, unlike the case with subcutaneous fat thickness, only with marked reductions in size are the parameters substantially affected. Also, unlike with subcutaneous fat thickness, the phase appears least affected rather than reactance.