Table 3.

Inter-day accuracy and precision of the LLOQ, LQC, MQC, HQC, and ULOQ evaluation samples for the rabbit whole blood method. Mean values in the table are average of the back-calculated concentrations from the standard curve.

QC sample (ng/mL) Compound A Metabolite
LLOQ LQC MQC HQC ULOQ LLOQ LQC MQC HQC ULOQ
4.05 10.5 131 1,640 2,030 4.04 10.5 132 1,650 2,020
n 18 18 18 18 18 18 18 18 18 18
Mean 4.30 10.5 127 1,630 2,080 4.19 10.9 130 1,620 1,960
CV (%) 4.2 2.5 3.4 3.5 2.7 6.3 4.8 2.3 2.0 1.8
Bias (%) 6.2 0.0 5.0 -0.6 2.5 3.7 3.8 -1.5 -1.8 -3.0