Figure 1.

Figure 1

A: Aortic root pressure waveforms from the same animal demonstrated suprasystolic diastolic augmentation and afterload reduction during 1:2 (middle panel) and 1:1 (bottom panel) intraaortic balloon pump (IABP) support compared to baseline (top panel). Arrows indicate supported beats. B: Diastolic augmentation of greater than 20 mmHg was achieved when the IABP was timed with a low-fidelity (fluid-filled) catheter in the radial artery or a high-fidelity (micromanometer) catheter in the aorta. AoP, Aortic Pressure; LV, left ventricular; *p<0.0001 vs. baseline; †p<0.0001 vs. LV failure.