Published in final edited form as: J Pharm Biomed Anal. 2013 Apr 16;83:1–9. doi: 10.1016/j.jpba.2013.04.007

Table 1.

Comparison of inter-day and intra-day imprecision and accuracy for plasma (over 21 days), DBS (manual), CSF and automated DBS extraction (all over three days).

Imprecision [RSD%] Accuracy [% of Nominal]
Plasma DBS CSF Plasma DBS CSF

Concentration
Intra-day:
198 ng/ml 4.8 6.7 8.3 93.5 102.0 110.0
593 ng/ml 2.7 4.5 9.2 104.0 99.0 104.0
16000 ng/ml 3.4 4.0 4.0 106.0 105.0 89.5
Inter-day: 198 ng/ml 8.7 8.3 11.7 99.1 103.0 102.0
593 ng/ml 3.4 6.3 10.9 107.0 104.0 100.0
16000 ng/ml 6.3 8.5 6.3 102.0 105.0 93.0
Automated Extraction
Intra-day: 198 ng/ml 17.8 94.9
593 ng/ml 5.8 96.5
16000 ng/ml 5.4 95.9
Inter-day: 198 ng/ml 8.1 96.4
593 ng/ml 15.3 90.2
16000 ng/ml 15.9 96.7