. 2014 May;52(5):1471–1477. doi: 10.1128/JCM.03313-13

TABLE 4.

Comparison of LAMP to microscopy for malaria diagnosis (n = 680 blood samples)

Method and result No. of results for microscopy
Sensitivity, % (95% CI) Specificity, % (95% CI) PPV, % (95% CI) NPV, % (95% CI) Diagnostic accuracy (%) Agreement (%)
Positive Negative
Microscopy
    Positive 199 0
    Negative 0 481 100 100 100 100 100 100
LAMP
    Positive 177 15
    Negative 22 466 88.9 (83.4–94.5) 96.9 (94.9–98.9) 92.2 (87.3–97) 95.5 (93.1–97.8) 94.6 86.7