Table 2.
MR scan #1 pre-MR |
MR scan #1 post-MR |
p Value | MR scan #2 pre-MR |
MR scan #2 post-MR |
p Value | MR scan #3 pre-MR |
MR scan #3 post-MR |
p Value | |
---|---|---|---|---|---|---|---|---|---|
Pacing capture threshold (V) | 1.99±1.12 | 2.00±1.11 | 0.97 | 1.91±1.14 | 1.92±1.12 | 0.88 | 1.74±1.2 | 1.78±1.2 | 0.85 |
Pacing lead impedance (ohm) | 528±95 | 507±84 | 0.45 | 527±102 | 538±88 | 0.73 | 524±89 | 520±88 | 0.97 |
High voltage lead impedance (ohm) | 50±8.3 | 51±8.4 | 0.68 | 52±9.2 | 53±6.1 | 0.79 | 51±7.9 | 51±6.3 | 0.97 |
Battery voltage (V) | 3.05±0.22 | 3.04±0.21 | 0.61 | 3.03±0.23 | 3.03±0.23 | 0.75 | 3.01±0.24 | 3.00±0.24 | 0.65 |
Values are presented as mean±SD.
Mann-Whitney two-sided p values.
ICD, implantable cardioverter defibrillator.