Table 2.

Difference in ICD interrogation data pre-and post-MR scan

MR scan #1
pre-MR
MR scan #1
post-MR
p Value MR scan #2
pre-MR
MR scan #2
post-MR
p Value MR scan #3
pre-MR
MR scan #3
post-MR
p Value
Pacing capture threshold (V) 1.99±1.12 2.00±1.11 0.97 1.91±1.14 1.92±1.12 0.88 1.74±1.2 1.78±1.2 0.85
Pacing lead impedance (ohm) 528±95 507±84 0.45 527±102 538±88 0.73 524±89 520±88 0.97
High voltage lead impedance (ohm) 50±8.3 51±8.4 0.68 52±9.2 53±6.1 0.79 51±7.9 51±6.3 0.97
Battery voltage (V) 3.05±0.22 3.04±0.21 0.61 3.03±0.23 3.03±0.23 0.75 3.01±0.24 3.00±0.24 0.65

Values are presented as mean±SD.

Mann-Whitney two-sided p values.

ICD, implantable cardioverter defibrillator.