Skip to main content
. 2014 Apr 23;26(4):557–562. doi: 10.1589/jpts.26.557

Table 2. Maximum power (watts) emitted by shortwave diathermy equipment considering the different simulated electrode-skin distances by equipment (n = 23).

Equipment Simulated electrode-skin distances (cm)

0.5 1.0 1.5 2.0 2.5 3.0
1 163 176 173 173 120 71
2 98 147.5 150 154 121 138.5
3 102 141 144 84 30 0
4 118 138 143 47 0 0
5 97 135 135 76 0 0
6 64 132.5 124 110 68 45
7 75.5 122.5 121 105 104 71.5
8 45 120 126 119.5 92.5 49
9 75.5 115.5 104 63 25 25
10 54 98 91.5 91 70.5 42
11 89 95 97 95 92.5 95.5
12 74 81 73.5 18 0 0
13 76 78.5 79 84 91 90
14 30 78 84 23 0 0
15 46 76.5 81 76 75 58
16 64 71.5 72 75.5 91.5 95
17 72 69 71 40 25 0
18 52 69 61 66 65 66
19 60 65 78 70 45 0
20 14 54 56 13 12 42
21 32.5 48 46 35 30 0
22 46 46 41.6 40 32.5 32.5
23 47 40 37.5 32.5 30 25