Table 3.

Computed calibration values for the body size and radiation dose conditions investigated in this study.

Phantom size Slope, Intercept, Slope, Intercept, CT# bone, CT# bone, CT# red marrow, CT# red marrow, CT# fat, CT# fat,
and dose level 80 kVp 80 kVp 140 kVp 140 kVp 80 kVp 140 kVp 80 kVp 140 kVp 80 kVp 140 kVp
Internal calibration
Small
0.95 125.32 0.63 109.35 1958.0 1310.5 124.7 109.5 −112.2 −88.3
0.95 124.24 0.63 109.35 1957.4 1311.8 123.6 109.8 −112.6 −87.6
0.96 124.84 0.63 109.52 1962.6 1312.5 124.5 109.9 −112.0 −87.2
Medium
0.93 127.02 0.61 109.46 1907.8 1283.2 127.0 109.9 −107.6 −85.8
0.93 124.94 0.61 109.53 1917.5 1278.1 124.9 109.5 −109.3 −86.1
0.94 125.62 0.61 109.22 1926.4 1284.5 125.5 109.2 −109.9 −86.1
Large
0.88 116.37 0.59 107.58 1805.7 1235.2 117.0 107.1 −106.9 −84.9
0.89 120.47 0.59 106.38 1831.4 1245.1 120.6 106.6 −109.2 −85.6
0.90 120.51 0.59 106.92 1857.1 1241.8 120.8 107.0 −109.2 −85.8
External calibration
Small, 4× 0.99 110.3 0.65 105.7 2013.9 1354.9 109.7 105.5 −131.3 −97.2
Medium, 4× 0.96 110.0 0.63 104.5 1954.7 1321.7 111.4 106.0 −132.3 −97.0
Large, 4× 0.94 108.7 0.62 101.4 1917.5 1293.5 108.0 102.1 −128.8 −97.0