Table 2. Interdevice comparison of central corneal thickness measurements.
Pairwise comparison | Mean paired difference |
95% CI of mean difference (µm) |
Pearson correlation |
|||
1mean±SD (µm) | 2P | Lower | Upper | R | 3P | |
OLCR and SD-OCT | 4.38±3.79 | <0.001 | 3.30 | 5.46 | 0.995 | <0.001 |
OLCR and Scheimpflug-Placido topographer | 4.38±6.03 | <0.001 | 2.66 | 6.09 | 0.985 | <0.001 |
USP and OLCR | 13.20±6.46 | <0.001 | 11.36 | 15.04 | 0.985 | <0.001 |
SD-OCT and Scheimpflug-Placido topographer | -0.05±6.77 | 0.987 | -1.93 | 1.92 | 0.980 | <0.001 |
USP and SD-OCT | 17.59±6.76 | <0.001 | 15.67 | 19.51 | 0.985 | <0.001 |
USP and Scheimpflug-Placido topographer | 17.58±8.13 | <0.001 | 15.27 | 19.89 | 0.977 | <0.001 |
1Mean interdevice difference; OLCR: Optical low coherence reflectometry; SD-OCT: Spectral-domain optical coherence tomography; USP: Ultrasound pachymetry; 2Repeated-measures ANOVA using Bonferroni adjustment for multiple comparisons; 3Pearson correlation analysis.