Skip to main content
. Author manuscript; available in PMC: 2015 May 1.
Published in final edited form as: Surv Ophthalmol. 2014 Jan 23;59(3):263–285. doi: 10.1016/j.survophthal.2013.09.002

TABLE 2.

In vivo confocal microscopy of corneal nerves: a comparison of imaging modalities

Type of Imaging Principal of Image Capture Aperture (μm) Depth of field (μm) Anterior stroma and subbasal nerve resolution* Measurement of subbasal nerve density (mm/mm2) Epithelial nerve imaging
In vivo confocal microscopy Tandem scanning 30.065 7.0 –11.065,210 Poor resolution (nerves)
High quality (stroma) 102,174
5.5 – 8.4 ± 2.064,65 No 65,156
Slit-scan (e.g., Confo-Scan) Variable: e.g., 300.065 10.0 – 26.065,102 High quality (nerves) 13
Poor resolution (stroma) 101,102
11.1 – 14.765,183 Possible, but unreliable 156
Laser (e.g., HRT-RCM) Variable 4.0 – 7.065,210 High quality (nerves) 180 21.7181 No 65,201

HRT-RCM = Heidelberg Retina Tomograph with Rostock Cornea Module.

*

Statements regarding resolution have been extrapolated from several articles. Oliveira-Soto et al. 174 observed 57.1% of subbasal plexus nerve images obtained with a tandem scanning microscope were blurred with heterogenous reflectivity. Simultaneously, 50% of nerves in the subepithelial plexus of the anterior stroma exhibited beads, suggesting higher quality resolution in the stroma. Tandem scanning also offers good repeatability when measuring thickness of corneal layers. 102 Slit-scanning confocal microscopy demonstrates subbasal plexus resolution at values greater than or equal to histologic analysis. 102 However, poor repeatability of corneal layer thickness measurements suggest inadequate stromal resolution. 101,102 Laser scanning confocal microscopy exhibits high quality resolution with increased visibility of finer subbasal nerve branches. 180 Differences in nerve resolution may also be supported by the variable measurements of subbasal nerve density.