. 2014 May 8;76:34–38. doi: 10.1016/j.curtheres.2013.11.005

Table I.

Temperature readings* and subjective evaluation by nurses, parents, and children related to the child’s comfort or ease of use.

Temperature measurement device Median Range
Rectal measurement 37.1 35.1–39.5
Tympanic thermometer 36.6 34.7–39.5
Infrared skin scan thermometer 36.7 35.0–39.7
Temporal arterial scan thermometer 37.1 35.6–39.8
Nurse Parent Child
Subjective evaluation, child friendly
 Rectal measurement 2 (0-5) 2 (0-5) 2 (0-5)
 Tympanic thermometer 1 (0-3) 1 (0-5) 1 (0-5)
 Infrared skin scan thermometer 1 (0-5) 1 (0-5) 1 (0-5)
 Temporal arterial scan thermometer 1 (0-5) 1 (0-5) 1 (0-5)
Subjective evaluation, ease of use
 Rectal measurement 3 (0-7)
 Tympanic thermometer 2 (0-6)
 Infrared skin scan thermometer 1 (0-9)
 Temporal arterial scan thermometer 2 (0-6)

Values are given in °C.

Based on Likert score of 0–5. Values are given as median (range).

Based on visual analog scale score of 0–10. Values are given as median (range).