Table 1.

Comparison of dose statistics between the treatment plans

Organs Dose statistic Breast case
Chest wall case
Isocentric plan
Nonisocentric plan Isocentric plan
Nonisocentric plan
Opposed tangential VMAT (2 arcs) Opposed tangential VMAT (2 arcs)
PTV prescription or mean dose: 50 Gy gEUD (Gy) 49.1 48.3 48.8 47.3 49.7 48.9
D95 (Gy) 36.6 32.2 38.2 40.9 42.9 44.1
D5 (Gy) 54.8 59.7 56.6 53.5 53.9 55.0
Uniformity index 1.50 1.85 1.48 1.31 1.26 1.25
Ipsilateral lung Mean dose or gEUD (Gy) 6.4 18.9 4.0 8.2 10.4 4.1
V5 (%) 20.8 89.5 16.6 25.5 74.1 18.1
V20 (%) 11.8 37.2 5.6 15.1 13.6 4.9
Heart Mean dose or gEUD (Gy) 0.9 26.0 0.8 0.4 8.9 0.4

Abbreviations: gEUD = generalized equivalent uniform dose; VMAT = volumetric modulated arc therapy.