Table 2.

Agreement of dose distribution in the phantom between GRD measurement and calculation

Range of dose ratio (calulation/measurement)
Without regression With regression
Test 1 88–162% 87–117%
Test 2 87–152% 79–121%
Test 3 89–137% 87–124%
Test 4 74–109% 78–111%