Table 3. Bioequivalence analysis of 32‐G × 4 mm vs 32‐G × 6 mm and 31‐G × 8 mm needles for insulin pharmacokinetics.

Ratioa 90% CI Lower/upper limitsb
Tmax (min)
32G × 4 vs 32G × 6 −0.3667 −5.8597, 5.1264 −7.9517, 7.9517
32G × 4 vs 31G × 8 −3.6167 −9.6907, 2.4574 −8.6017, 8.6017
Cmax (μU/mL)
32G × 4 vs 32G × 6 −0.0109 −0.1713, 0.1496 −0.2230, 0.2231
32G × 4 vs 31G × 8 −0.0164 −0.1893, 0.1565 −0.2230, 0.2231
AUC0–∞ (min μU/mL)
32G × 4 vs 32G × 6 −0.0608 −0.1827, 0.0611 −0.2230, 0.2231
32G × 4 vs 31G × 8 −0.0810 −0.2076, 0.0457 −0.2230, 0.2231

CI, confidence interval; Tmax, time to the maximum concentration; Cmax, maximum concentration; AUC0–∞, area under the curve from 0 to infinity min.

a

Ratio of the mean values.

b

±20% for time to the maximum concentration (Tmax), log (0.80), log (1.25) for maximum concentration (Cmax) and area under the curve from 0 to infinity min (AUC0–∞).