Skip to main content
. Author manuscript; available in PMC: 2014 Jun 1.
Published in final edited form as: Clin Chem. 2013 Feb 8;59(6):982–990. doi: 10.1373/clinchem.2012.195594

Table 1.

Within-run, between-run and total imprecision of LC-MS/MS method for thyroglobulin.

Sample Concentration,
ng/mL
Within-run,
%
Between-
run/day, %
Total, %
Low 1# 2.1 6.75 3.67 7.69
Low QC 1* 2.3 na 13.9 na
Low 2# 5.7 6.87 5.96 9.10
Medium QC 2* 6.5 na 10.5 na
Medium# 14.8 6.56 5.40 8.50
High# 399 3.56 1.71 3.95
High QC* 172.8 na 3.5 na
#

Samples analyzed in three replicates per day over five days

*

Samples analyzed in one replicate per day, over 20 days