TABLE 3.

Summary of Data on the Pediatric Neuroradiological Cases Used for Dose Calculations

Age, years (reference age for specifying phantom) Number of cases Weight (kg) Height (cm) MSDa, mGy (range) Combined fluoroscopy time, minutes (range) Total DAP, Gy·cm2 for PA + LAT (range) Percentage DAP from frontal plane (range)
<1 (0) 3 4.5 (1.8–3.6) 48 (30–64) 910 (380–1400) 69 (50–94) 5.9 (4.3–8.1) 53 (46–60)
1–2 (1) 12 10.5 (7.7–14.5) 75 (70–84) 910(240–1810) 42 (15–132) 11.4 (2.5–28.2) 59 (28–78)
3–7 (5) 10 20.5 (14.5–29.5) 110 (97–122) 1640 (400–3190) 55 (13–132) 14.3 (2.9–28.5) 66 (51–81)
8–12 (10) 5 42.3 (34–51.3) 150 (137–163) 2370 (1500–3810) 98 (37–138) 27.8 (13.9–47.4) 59 (54–65)
13–17 (15) 19 56.6 (20–88.4) 163 (122–180) 2290 (600–6610) 110 (44–314) 31.3 (7.2–89.8) 63 (47–100)
Overall 49 33 (1.8–88.4) 122 (30–180) 1740 (240–6610) 79 (13–314) 20.8 (2.5–89.8) 62 (28–100)
a

Maximum skin dose.