Table 2.
Patient characteristics | |||||
• More than 6 weeks after pacemaker implantation | |||||
• Right or left pectoral implant (prepectoral or submuscular) | |||||
• No other contraindications to MRI (eg, non-MRI-conditional abandoned leads) | |||||
• No lead adapters or extenders in situ | |||||
MRI scanner and scanning parameters | |||||
• Static magnetic field of 1.5 T | |||||
• Cylindrical bore | |||||
• Maximum SAR of 2 W/kg | |||||
• Maximum head SAR of 3.2 W/kg | |||||
• Maximum gradient slew rate of 200 T/m per second | |||||
Pacemaker model | |||||
Medtronic® Revo™, Ensura™ and Advisa™ Sure Scan® pacemakers with CapSure™ leads | Biotronik® Evia™, Entovis™, Estella™ and Ecuro™ ProMRI® pacemakers with Safio™ and Solia™ leads | St Jude® Accent™ MRI pacemaker with Tendril™ leads | Boston Scientific® Ingenio™ and Advantio™ Image Ready® pacemakers with FINELINE™ leads | Sorin® KORA 100™ SR and DR pacemakers MRI with BEFLEX™ leads | |
MRI-conditional modes | All | ||||
Stable pacing capture threshold values | ✓ | ||||
Capture threshold and pulse width | ≤2.0 V @ 0.4 msec | ≤2.0 V @ 0.4 msec | ≤2.5 V @ 0.5 msec | ≤3.0 V @ 0.4 msec | ≤3.0 V @ 0.5 msec |
Lead impedance 200–1,500 Ohms | All | ||||
Battery charging status at least 30% | ✓ | ||||
No diaphragmatic pacing at 5.0 V at a pulse width of 1.0 msec | ✓ | ✓ | |||
Scan zone restrictions | None | Exclusion zone between C1 and T12 vertebrae | None | None | Exclusion zone between C1 and T12 vertebrae |
Scan duration restrictions | None | ≤30 minutes for each scan; total maximum of 10 hours | None | None | None |
Abbreviations: SAR, specific absorption rate; MRI, magnetic resonance imaging.