Table 7.
Instrument | NOVA View | AKLIDES | EUROPattern | Image Navigator | Helios | ZENIT G Sight |
---|---|---|---|---|---|---|
Manufacturer | INOVA diagnostics | Medipan | Euroimmun | Immunoconcepts | Aesku | Menarini |
| ||||||
LIMS connection (software) | Yes (QUANTA Link) |
Yes (system independent, standard XML interface) | Yes (EUROLabOffice) | Yes (direct) Optional: lab traffic control |
Yes (direct) Optional: Aesku.Lab middleware |
Yes (ZenIT) |
| ||||||
Slide identification via barcode | Yes by handheld scanner | Yes by handheld scanner | Yes by integrated scanner | Yes | Yes by integrated scanner | Yes by integrated scanner |
| ||||||
Loading capacity | 5 slides (up to 60 wells) |
5 slides (up to 60 wells) |
50 slides (up to 500 wells) |
4 slides (up to 84 wells) |
20 slides (up to 240 wells) |
5 slides (up to 70 wells) |
| ||||||
Image acquisition speed | ~45 s/well for 3 images | ~40 s/well | <20 s/well | ~25 s/well for 4 images | 10 s/picture Customizable from 1 to 10 images | >60 s/well number of pictures: 5 (small scan), 50 (medium scan), or 220 (full scan) |
| ||||||
100% QC for substrate and process integrity/counterstaining | Yes/DAPI | Yes/DAPI | Yes/Propidium iodide | None/None | None/None | None/None |
| ||||||
Automatic pos./neg. discrimination incl. presorting of images | Yes | Yes | Yes | Yes | Yes | Yes |
| ||||||
Batchwise verification of negative samples | Yes | Yes | Yes | Yes | Yes | Yes |
| ||||||
Automatic pattern recognition | Yes | Yes | Yes | No | No | Yes |
| ||||||
Pattern Analysis method | Pattern recognition by mathematical algorithm | Pattern recognition by mathematical algorithm | Pattern recognition by mathematical algorithm | No pattern matching capabilities | No pattern matching capabilities | Pattern recognition by mathematical algorithm |
| ||||||
Number of recognizable ANA staining pattern list out | 6 Homogeneous Speckled Centromere Nucleolar Nuclear dot Cytoplasm Negative Positive unrecognized |
10 Homogeneous Speckled Centromere Nucleolar Nuclear dot Cytoplasm Negative Positive unrecognized |
8 Homogeneous Speckled Centromere Nucleolar Nuclear dot Cytoplasm Negative Nuclear rim Mixed pattern mitotic |
None | None | 5 Homogeneous Speckled Centromere Nucleolar Cytoplasm Negative |
| ||||||
Analysis of mixed staining pattern | Limited (homogeneous/ nucleolar) |
Yes | Yes | No | No | No |
| ||||||
Merged results per patient (different dilutions) | Yes | Yes | Yes | Yes | Yes | Yes |
| ||||||
Final result validation possible while system processes remaining samples | Yes | No | Yes | Yes | Yes | No |
| ||||||
Instrument calibration to minimize variability | Yes | Yes | Yes | Yes | Yes | No |
| ||||||
Integration with slide processing in 1 instrument | No | No | No | No | Yes | No |