Skip to main content
. 2014 Apr 27;2014:315179. doi: 10.1155/2014/315179

Table 7.

Automated digital ANA reading systems.

Instrument NOVA View AKLIDES EUROPattern Image Navigator Helios ZENIT G Sight
Manufacturer INOVA diagnostics Medipan Euroimmun Immunoconcepts Aesku Menarini

LIMS connection (software) Yes
(QUANTA Link)
Yes (system independent, standard XML interface) Yes (EUROLabOffice) Yes (direct)
Optional: lab traffic control
Yes (direct)
Optional: Aesku.Lab middleware
Yes (ZenIT)

Slide identification via barcode Yes by handheld scanner Yes by handheld scanner Yes by integrated scanner Yes Yes by integrated scanner Yes by integrated scanner

Loading capacity 5 slides
(up to 60 wells)
5 slides
(up to 60 wells)
50 slides
(up to 500 wells)
4 slides
(up to 84 wells)
20 slides
(up to 240 wells)
5 slides
(up to 70 wells)

Image acquisition speed ~45 s/well for 3 images ~40 s/well <20 s/well ~25 s/well for 4 images 10 s/picture Customizable from 1 to 10 images >60 s/well
number of pictures: 5 (small scan), 50 (medium scan), or 220 (full scan)

100% QC for substrate and process integrity/counterstaining Yes/DAPI Yes/DAPI Yes/Propidium iodide None/None None/None None/None

Automatic pos./neg. discrimination incl. presorting of images Yes Yes Yes Yes Yes Yes

Batchwise verification of negative samples Yes Yes Yes Yes Yes Yes

Automatic pattern recognition Yes Yes Yes No No Yes

Pattern Analysis method Pattern recognition by mathematical algorithm Pattern recognition by mathematical algorithm Pattern recognition by mathematical algorithm No pattern matching capabilities No pattern matching capabilities Pattern recognition by mathematical algorithm

Number of recognizable ANA staining pattern list out 6
Homogeneous
Speckled
Centromere
Nucleolar
Nuclear dot
Cytoplasm
Negative
Positive unrecognized
10
Homogeneous
Speckled
Centromere
Nucleolar
Nuclear dot
Cytoplasm
Negative
Positive unrecognized
8
Homogeneous
Speckled
Centromere
Nucleolar
Nuclear dot
Cytoplasm
Negative
Nuclear rim
Mixed pattern
mitotic
None None 5
Homogeneous
Speckled
Centromere
Nucleolar
Cytoplasm
Negative

Analysis of mixed staining pattern Limited (homogeneous/
nucleolar)
Yes Yes No No No

Merged results per patient (different dilutions) Yes Yes Yes Yes Yes Yes

Final result validation possible while system processes remaining samples Yes No Yes Yes Yes No

Instrument calibration to minimize variability Yes Yes Yes Yes Yes No

Integration with slide processing in 1 instrument No No No No Yes No