Table 2.

Comparison of detection of respiratory virus by three automatic extraction methods from nasopharyngeal swab specimens.

Virus QIAcube EZ1 advanced XL MICROLAB Nimbus IVD Total number of specimens
RV + + + 14
+ + 3
+ 1a
+ + 1
+ 2b

MPV + + + 7
+ + 1

PIV3 + + + 8

PIV1 + + + 1
+ 1b

PIV4 + 1b

RSV-A + + + 7
+ 1b

ADV + + + 2
+ + 1
+ + 2
+ 3b

FluB + + + 2
+ 1a

FluA + + + 2

BoV + + + 1
+ 1a

EV + + + 2
+ 1b

CoV-OC43 1 1 1 1

CoV-NL63 + 1b

CoV-229E + 1b

Total 57 59 54 68

aVirus detected by only 1 of the three automated nucleic acid extraction systems, and we confirmed this result by sequencing analysis and confirmed it as “true positive.”

bVirus detected by only 1 of the three automated nucleic acid extraction systems, but they were not detected by repeated test with the Seeplex RV15 ACE Detection kit (Seegene).