Skip to main content
. 2014 Apr 28;2014:430650. doi: 10.1155/2014/430650

Table 3.

Comparison of the PCR results from 3 automated nucleic acid extraction systems using a total of 84 nasopharyngeal specimens.

QIAcube EZ1 advanced XL MICROLAB Nimbus IVD
Positive results 45 (53.6%) 45 (53.6%) 43 (51.2%)
 True positivea 42 38 41
 False positivea (virus type) 3 (PIV4, EV, RSV-B) 7 (PIV1, ADV (n = 3), RSV-A, RV (n = 2)) 2 (CoV-NL63, CoV-229E)
Negative results 39 (46.4%) 39 (46.4%) 41 (48.8%)
 True negativea 36 34 35
 False negativea (virus type) 3 (FluB, RV, ADV) 5 (BoV, RV (n = 4)) 6 (FluB, RV, MPV, 
BoV, ADV (n = 2))
Sensitivity (%) 93.3 88.4 87.2
Specificity (%) 92.3 82.9 94.6
Concordance rate (%) 94.2 88.3 92.2
Kappa coefficientb (95% CI) 0.88 (0.79 to 0.97) 0.76 (0.64 to 0.89) 0.84 (0.74 to 0.95)
RV sensitivity/specificity 94.7/100.0 78.9/96.9 94.7/100.0
MPV sensitivity/specificity 100.0/100.0 100.0/100.0 85.0/100.0
PIV3 sensitivity/specificity 100.0/100.0 100.0/100.0 100.0/100.0
RSV-A sensitivity/specificity 100.0/100.0 100.0/98.7 100.0/100.0
ADV sensitivity/specificity 80.0/100.0 100.0/96.2 60.0/100.0
Characteristics of three systems
 Principle Spin column Magnetic particle Magnetic particle
 Sample volume (μL) ≤200 100–400 600
 Elution volume (μL) 20–150 60–150 100
 Turnaround timec (min) 90 45 150
 Sample capacity 12 14 48
 Automated PCR set-up Impossible Impossible Possible

aWhen all three automated nucleic acid extraction systems yielded the same results, they were considered “true positive” or “true negative.” If only 1 system yielded negative result, it was considered “false negative.” When the pathogen was detected from only 1 system, multiplex PCR and sequencing analysis were performed to confirm the result. b P < 0.0001 for all values. cincludes lysis step but not the hands-on time.