Skip to main content
. 2014 May 1;16(5):2419. doi: 10.1007/s11051-014-2419-x

Table 1.

Quantitative validation of NanoTrackJ (NTJ) using the polystyrene experiments shown in Fig. 1

Polystyrene diameter 100 nm 100 nm 200 nm 200 nm
Software NTA NTJ NTA NTJ
Analysis time (s) 60 60 60 60
Modal value (nm) 101 98 181 196
IQR75-25 (nm) 22 24 31 31
#Tracks 1,263 1,290 524 469
Tracklength 53 64 80 109
Figure 1a 1b 1c 1d