Standard microcatheters in use.

  Finecross (Terumo) Supercross (Vascular perspectives) Valet (Volcano) Corsair (Asahi) Tornus (Asahi)
Composition/design -Stainless steel braid structure
-Hydrophilic coating for distal 60cm
-PTFE inner layer
-Stainless steel braid structure
-Hydrophilic coating for distal 40cm
-PTFE inner layer
-45°, 90°, 120° angled tip versions
-Stainless steel braid structure with variable metal wind thickness and powder plastic coating
-Hydrophilic coating distal 30cm
-Shapeable distal tip 
-Tungsten braiding with 10 elliptical steel braids
-Soft, tapered tip
-Hydrophilic polymer coating for distal 60cm
-Inner polymer lumen
-Eight individual stainless steel wires, braided with a left-handed thread
Diameter/ length -Tapers from 2.6Fr to 1.8Fr over entire catheter
-Available 130 and 150cm length
-Tapers from 2.5Fr to 1.8Fr
-Available 130 and 150cm length
-Available 2 diameters (1.8Fr and 3.5Fr)
- Available 130 and 150cm length
-2.8Fr with tapered distal tip (0.016 inch)
-Available 135 (antegrade) and 150cm (retrograde) length
-Available in two sizes: Tornus (2.1Fr and 2.6Fr)
-135cm length
Instructions for use -Rotation in either direction facilitates forward motion -Rotation in either direction facilitates forward motion -Rotation in either direction facilitates forward motion -Rapid rotation in either direction enhances forward motion -Rotated into lesion counter-clockwise (maximum 20 turns in one go) and withdrawn clockwise
-Should not be flushed (porous design and can splay wires)
Indication -Provision of support for guidewire crossing -Provision of support for guidewire crossing -Provision of support for guidewire crossing -Provision of support for guidewire crossing -Collateral channel crossing (especially retrograde) -Highly supportive catheter
-Allows balloon tracking (once distal wire position achieved when other microcatheters or OTW balloons will not cross) or affords rotawire exchange and therefore useful combination with rotablation