pH values measured with the new electronic device and the pH-meter in fresh urine

pH interval pH-meter Mean (±SD) New device Mean (±SD) Difference (±SD) n
5-5.5 5,17 ± 0.08 5,20 ± 0.12 0.03 ± 0.14 10
5.5-6 5,60 ± 0.14 5,68 ± 0.14 0.08 ± 0.20 10
6-6.5 6,26 ± 0.14 6,29 ± 0.12 0.03 ± 0.18 7