Table 2.
Number of scans affected | Description | Total Exposed (FAS, N = 37) | Total Controls (N = 15) |
---|---|---|---|
5 | Phase unwrapping algorithm generated a 2π phase offset, resulting in a frequency shift of 208 Hz. This rendered water suppression ineffective. | 3 | 2 |
4 | Ghosting in the first vNav EPI volume rendered the vNav unusable. This resulted from large shim terms set by the scanner for the voxel. | 3 | 1 |