. 2014 Apr 30;4(5):e004565. doi: 10.1136/bmjopen-2013-004565

TableĀ 3.

Diagnostic performance of Omron and Merlin single-lead ECGs for detecting atrial fibrillation

n=999 Prevalence 7.9% Interpretation Omron single-lead ECG (95% CI) Merlin single-lead ECG (95% CI)
Sensitivity (%) Cardiologist: 1 95 (82 to 99) 97 (85 to 100)
2 95 (82 to 99) 91 (77 to 98)
3 90 (77 to 97) 98 (87 to 100)
4 98 (87 to 100) 88 (74 to 96)
Specificity (%) Cardiologist: 1 75 (71 to 79) 76 (71 to 80)
2 98 (96 to 99) 95 (93 to 97)
3 97 (96 to 99) 92 (89 to 94)
4 95 (92 to 97) 91 (88 to 94)
Positive likelihood ratio Cardiologist: 1 3.8 (3.2 to 4.6) 3.99 (3.3 to 4.8)
2 41.8 (22.5 to 77.5) 18.8 (12.2 to 28.8)
3 35.8 (20.3 to 63.1) 11.5 (8.5 to 15.6)
4 18.5 (12.6 to 27.2) 10.1 (7.4 to 13.9)
Negative likelihood ratio Cardiologist: 1 0.07 (0.02 to 0.28) 0.04 (0.01 to 0.26)
2 0.05 (0.01 to 0.21) 0.09 (0.03 to 0.27)
3 0.098 (0.04 to 0.25) 0.03 (0.004 to 0.18)
4 0.02 (0.004 to 0.17) 0.13 (0.06 to 0.3)
Positive predictive value Cardiologist: 1 24.1 (17.4 to 31.9) 24.5 (17.6 to 32.5)
2 77.8 (62.9 to 88.8) 60.4 (46.0 to 73.5)
3 76.0 (61.8 to 86.9) 50.6 (39.3 to 61.9)
4 62.1 (49.3 to 73.8) 47.4 (36.0 to 59.1)
Negative predictive value Cardiologist: 1 99.4 (97.9 to 99.9) 99.7 (98.3 to 100)
2 99.5 (98.3 to 99.9) 99.3 (97.9 to 99.8)
3 99.1 (97.8 to 99.8) 99.8 (98.7 to 100)
4 99.8 (98.8 to 100) 98.9 (97.3 to 99.6)