Abstract
Background
The electrocardiographic (ECG) Tpeak-Tend interval (TpTe) is associated with arrhythmias and sudden cardiac death. TpTe offers a supplementary measure for the QT interval (QT), yet its repeatability has not been established.
Purpose
Evaluate short-term repeatability of TpTe and QT.
Methods
Four ECGs were obtained on sixty participants. The sources of variation, intra-class correlation coefficient (ICC) - an index of reproducibility, and minimal detectable change (MDC) were estimated for TpTe and QT. The impact of repeated measurements on repeatability was estimated for a hypothetical clinical trial designed to detect drug-induced prolongation of TpTe and QT.
Results
We used heart rate-adjusted QT [(QT)a] but TpTe in the study group was rate-invariant. The ICC [95% confidence interval (CI)] was 0.77 (0.69, 0.85) for TpTe, 0.75 (0.65, 0.85) for QT and 0.68 (0.57, 0.79) for (QT)a. The MDC (ms) was 21, 32 and 26 for TpTe, QT and (QT)a respectively.
Conclusion
TpTe has excellent repeatability supporting its use as a supplement to QT in observational and clinical studies.
Keywords: Electrocardiogram, QT, TpTe interval, Repeat variation T
Introduction
The Tpeak-Tend interval (TpTe) from the resting standard 12-lead electrocardiogram (ECG) is a novel measure of myocardial repolarization, reported to be associated with incident ventricular arrhythmia, heart failure, and sudden cardiac death (SCD) (1-4). Prolongation of TpTe represents a period of potential vulnerability to reentrant arrhythmias offering a supplement to the traditional QT interval (QT) (5) in the setting of the established associations between QT prolongation with total, cardiovascular and sudden cardiac death (1). Although widely used to quantify ventricular repolarization abnormalities, QT has limitations and varying recommendations have been issued for an appropriate heart rate correction (6-8). An increased interest in other measures of myocardial repolarization has thus emerged, including TpTe (5). The peak of the T wave occurs when the largest number of left ventricular lateral wall myocytes is at phase 3 of their repolarization and the uncanceled repolarization wave front is largest according to the infinite-medium model of ventricular repolarization (8, 9). This assertion is supported by electrophysiological evidence and electrocardiographic potential theory applied to the generation of the T wave (10-12). Opthof et al. (13) have pointed out that the interval from T wave onset to Te reflects total dispersion of repolarization, therefore providing more comprehensive clinical information of arrhythmic risk.
The association of TpTe with ventricular arrhythmia and SCD and its consideration as a supplement for QT makes it pertinent to evaluate the repeatability of TpTe, which to our knowledge has not been addressed. The aim of our study was to characterize the repeatability and main sources of variability of TpTe, and to contrast these properties with those of QT.
Methods
Study participants and protocol
Study participants included sixty male and female volunteers free of cardiovascular symptoms recruited in Chapel Hill, NC. Details of study participant recruitment were described elsewhere (15). Briefly, efforts were made to match the age, gender and racial characteristics of the volunteer group to that of the Atherosclerosis Risk in Communities (ARIC) cohort at baseline and included participants between 45-64 years of age, free of diabetes, hypertension, emphysema, chronic obstructive pulmonary disease, congestive heart failure, renal disease, a pacemaker, and were not taking class I or III antiarrhythmics. Informed consent was obtained from all participants. This study was approved by the University of North Carolina (UNC) School of Medicine institutional review board. All examinations were conducted at the General Clinical Research Center of the UNC Hospitals.
Electrocardiographic methodology
The protocol used by the ARIC study was followed for electrode positioning, room condition, data collection and transfer (15). Following identical protocol at both study visits, two sets of ECGs were obtained on the same study participant at each visit. Four digital 12-lead ECGs (ECG1, ECG2, ECG3, and ECG4) were obtained by one of four trained and certified technicians at two different examination visits [two ECGs per visit, ten days apart on average, standard deviation (SD)=four days] both of which were conducted between 7:30 and 11:30 am, following a 12-hour period during which participants were asked not to consume anything other than water and to refrain from smoking and intense physical activity. The two ECG recordings at each study examination were separated by one to two minutes and were performed without removal of the electrodes. The trained and certified technicians positioned Kendall Q-Trace 5400 Ag/AgCl electrodes (Ludlow Co, Chicopee, MA) using the E-V6 half-point method and a standardized protocol for placing electrodes on women after a 15-minute supine rest (16). ECGs were recorded using a MAC Personal Cardiographer (Marquette Electronics, Inc., Jupiter, FL). ECGs were digitized at 250 samples per second per lead and sent via modem after each recording session to the Epidemiology Cardiology Research (EPICARE) Center at the Wake Forest University School of Medicine in Winston-Salem, NC. The EPICARE reading center, masked to participant identity and previous ECG readings, processed the ECGs with the Marquette program GE 12SL software 2001 version (GE, Milwaukee, WI). All ECGs were visually reviewed, and all measurements were validated using interactive software. Simultaneous ECGs were displayed to the reviewers expanded on the graphics screen with computer-placed global time lines at QRS onset, QRS end and Tend. The reviewers had the option either to accept computer time lines or to move them to a correct location if in error. The algorithms for detection of Tend in the GE 12SL program are based on combined slope and amplitude thresholds (17). Simultaneous TpTe was measured as the interval (ms) from the peak to the end of the global T wave (the T wave vector magnitude curve).
Statistical analysis
Our analysis excluded three participants from the original sample size of sixty three; two with extreme outliers (TpTe<50 ms and QT > 480 ms) between 2.5 and 4 SD away from the mean, and one for evidence of an arrhythmia indicated by frequent ventricular premature beats. We calculated the frequency and percent of selected categorical demographic variables and the mean and standard deviation of interval-scale measures. Linear formulas were used to calculate heart rate-adjusted QT as (QT)a =QTe +127*(1-RR) for men and (QT)a = QTe+136*(1-RR for women). Bazett rate-adjusted QT (QT)bz was calculated as [QTe/(RR) ½] (18). Mean values and SD for TpTe, (QT)a and (QT)bz at each study visit were calculated. We used random-effects models: Yijk = μ + TpTe + Vj(i) = ek(ij) where μ is the intercept, i=1, 2, 3 to the 60th participant, j is either visit 1 of visit 2 and k is the first or second ECG recording to estimate sources of variation in TpTe and QT [between-participant , between-visit , and within-visit ]. Two assumptions were that within-visit variation is represented by the residual variation in the model, between-participant variation was the same for all participants, and within-visit variation was the same for all visits and all participants, as described previously (6).
Intraclass correlation coefficient and standard error of measurement
The intraclass correlation coefficient (ICC) was defined as the ratio of between-participant variance to total variance observed in a repeated measurements study. where is the variance between participants, is the variance between visits, and is the variance within visit. Thus high ICC indicates that the variation of the parameter of interest is mainly due to variation between participants suggesting high repeatability within participants. Variance between technicians was negligible. We also calculated the standard error of measurement . The minimal detectable change (MDC = SEM*√2*1.96) was calculated with 95% confidence intervals (CI).
95 % Confidence intervals
We also estimated the impact of repeated measurements of TpTe and QT on the precision of the TpTe and QT measurements for a hypothetical clinical trial as a function of the number of measurements, n, and the sample size of such a trial, N. Since no previous requirement was issued for designing clinical trials powered to detect TpTe prolongation that would identify possible side effects of a drug, we used the requirement of 5-ms that was established for QT (6). This aspect of the repeatability analysis allows us to examine the ability to detect side effects associated with prolongation of TpTe and QT that may be encountered during a clinical trial especially those side effects resulting from drugs. The 95% CIs were calculated for the “true” change in the mean TpTe and QT intervals in a hypothetical clinical trial averaged over replicate measures and the SD (estimate of change ±1.96 SD).
The SD for this estimate of change was calculated the same way for QT and TpTe assuming a clinical trial with 2 arms where X’ was the mean postdrug-predrug change in TpTe was the mean postdrug-predrug change in TpTe in the treatment group, and X’T=0 was the mean postdrug-predrug change in TpTe in the control group. The variance (Var) of the difference in TpTe change between the treatment and control group was: Var (X’T=1- X’T=0) = Var (X’T=1) + Var(X’T=0). Based on the assumption that the variances in the treatment and control group are identical, the variance of the difference in TpTe change between the treatment and control group was calculated as the following: 4*within-participant variance/ (n*N) where n is the number of replicate measures and N is the sample size.
Similar to the calculations above that were for group level changes, the 95% CI for individual changes were calculated as the estimate of change ±1.96*√2*intra-individual SD/√n, where n is the number of repeated measurements. All data were analyzed using SAS version 9.3 (SAS Institute, Inc., Cary, NC).
Results
Characteristics of study participants
The demographic and clinical characteristics of study participants are shown in Table 1. Participants included equal numbers of males and females who mostly self-identified as white (68%), with a mean age of 52±5 years, and an average body mass index (BMI) of 27±5. Medication use was low; anticholinergic (8%), beta-blocker (2%), diuretic (6%), selective serotonin reuptake inhibitor (17%), sympathomimetic (7%), and thyroid medication (5%). A first degree atrioventricular block was observed in 15% of the study participants. Non-fasting/smoking twelve-hour pre-examination was rare at either visit, 7%.
Table 1.
Variable | Number (%) or Mean ± standard deviation |
---|---|
Female | 30 (50) |
Non-white | 19 (32) |
Age (years)* | 52±5 |
Body mass index (kg/m2) * | 27±5 |
Medication Use | |
Anticholinergic | 5 (8) |
Beta-Blocker | 1 (2) |
Diuretic | 4 (7) |
Selective serotonin reuptake inhibitor (SSRI) | 10 (17) |
Sympathomimetic | 4 (7) |
Thyroid medication | 3 (5) |
Atrioventricular block (per subject) | 9 (15) |
Non-fasting/smoking <12 hours before either visit | 4 (7) |
n=59
We reported (QT)a but TpTe in the study group was rate-invariant. The mean values and the standard deviations of TpTe were similar across all four measurements at the two visits. The mean values and the standard deviations of QT and (QT)a were also similar across all four measurements at the two visits with smaller values overall for (QT)a compared to QT (Table 2). The absolute between-visit difference was (7 ± 5 ms) for TpTe and was similar to the within-visit difference (8 ± 6 ms). The absolute value of the between-visit difference (7 ± 6 ms) for QT was also similar to the within-visit difference (6 ± 6 ms). The absolute value of the between-visit difference and the within-visit difference for (QT)a respectively were (9 ± 8 ms) and (7 ± 5 ms) (Table 2).
Table 2.
Measure (ms) | Visit 1 mean (SD)
|
Visit 2 mean (SD)
|
Absolute mean differences (SD), range
|
|||
---|---|---|---|---|---|---|
ECG1 N=59 |
ECG2 N=60 |
ECG3 N=60 |
ECG4 N=59 |
Between-visit N=59 |
Within-visit N=59 |
|
*TpTe | 90 (16) | 89 (15) | 89 (17) | 89 (17) | 7 (5), 1-30 | 8 (6), 1-30 |
QT | 417 (25) | 415 (23) | 415 (21) | 417 (22) | 7 (6), 0-34 | 6 (6), 0-31 |
‡(QT)a | 413 (16) | 415 (16) | 412 (15) | 415 (12) | 9 (8), 1-45 | 7 (5), 0-45 |
§(QT)bz | 412 (19) | 415 (20) | 410 (18) | 414 (14) | 8 (6), 1-40 | 10 (9), 0-51 |
TpTe =T peak (Tp) to T end (Te) interval (unadjusted)
QT=QT interval (unadjusted)
(QT)a =QTe+127*(1-RR) for men (QT)a=QTe+136*(1-RR) for women (rate-adjusted QTe)
(QT)bz=Rate-adjusted QT by Bazett’s formula [(QT/(RR)1/2]
Components of measurement error are listed in Table 3. Between-participant variation accounted for 77% of the total variation in TpTe, 75% of QT, 60% of (QT)a, and 69% of (QT)bz. Within-visit variation for TpTe accounted for a large proportion of the total variation than the between-visit variation (20% vs. 3%) in the random effects model. The within-visit variation for (QT)a. also accounted for more (27%) than the between-visit (13%) variation.
Table 3.
TpTe | QT | (QT)a | (QT)bz | |||||
---|---|---|---|---|---|---|---|---|
| ||||||||
Source of variation | SD (ms) | % total | SD (ms) | % total | SD (ms) | % total | SD (ms) | % total |
Between-participant | 14 | 77 | 20 | 75 | 11 | 60 | 16 | 69 |
Between-visit | 3 | 3 | 9 | 15 | 5 | 13 | 6 | 9 |
Within-visit | 7 | 20 | 7 | 10 | 8 | 27 | 9 | 21 |
Total | 16 | 100 | 23 | 100 | 15 | 100 | 19 | 100 |
ICC (95% CI) | 0.77 (0.69, 0.85) | 0.75 (0.65, 0.85) | 0.60 (0.47, 0.73) | 0.68 (0.57, 0.79) | ||||
SEM (ms) | 8 | 11 | 9 | 6 | ||||
MDC (ms) | 21 | 32 | 26 | 29 |
SD: standard deviation; ICC: intra-class correlation coefficient; CI: confidence interval; SEM: standard error of measurement; MDC: minimum detectable change.
ICC=(between-participant variance)/(total variance)
SEM=√(between-visit variance + within-visit variance)
MDC=SEM*√2+1.96
TpTe had an ICC of 0.77 and the MDC between two measures was 21 ms. The ICC for (QT)a and (QT)bz were 0.60 and 0.68 respectively. The MDC was 32 ms for (QT)a and 29 ms for QTbz (Table 3).
Table 4 shows the impact of repeated measurements of TpTe and QT on precision. The 95% CIs for a mean 5 ms effect on TpTe and QT was calculated as a function of the number of measurements (n) and the total sample size (N). The more precise CIs were seen when the number of measurements and number of participants was larger. For example, when two ECGs are available from baseline and there are 60 enrolled participants in a clinical trial, if the placebo-corrected increase was 5 ms, the 95% CI around this estimate for TpTe would be between 2 and 8 ms and between 1 and 9 ms for (QT)a. Calculations of these estimations at the individual level show much larger imprecision (Table 5). For example, when two ECGs are available from baseline for a participant, this estimate ranges between 5 and 35 ms for TpTe and between 2 and 38 ms for (QT)a (Table 5).
Table 4.
TpTe | QT | (QT)a | ||||||
---|---|---|---|---|---|---|---|---|
| ||||||||
N | n | 95% CI | N | n | 95% CI | N | n | 95%CI |
40 | 1 | 0,10 | 40 | 1 | -2, 12 | 40 | 1 | -1, 11 |
2 | 2, 8 | 2 | 0, 10 | 2 | 1,9 | |||
3 | 2, 8 | 3 | 1, 9 | 3 | 2,9 | |||
4 | 3, 7 | 4 | 1, 9 | 4 | 2,8 | |||
5 | 3,7 | 5 | 2, 8 | 5 | 2,8 | |||
6 | 3,7 | 6 | 2, 8 | 6 | 3,7 | |||
7 | 3, 7 | 7 | 2, 8 | 7 | 3,7 | |||
8 | 3, 7 | 8 | 2, 8 | 8 | 3,7 | |||
9 | 3, 7 | 9 | 3, 7 | 9 | 3,7 | |||
10 | 3, 7 | 10 | 3, 7 | 10 | 3,7 | |||
60 | 1 | 1, 9 | 60 | 1 | -1, 11 | 60 | 1 | 0, 10 |
2 | 2, 9 | 2 | 1, 9 | 2 | 2, 8 | |||
3 | 3, 7 | 3 | 2, 8 | 3 | 2, 8 | |||
4 | 3, 7 | 4 | 2, 8 | 4 | 3, 7 | |||
5 | 3, 7 | 5 | 2, 8 | 5 | 3, 7 | |||
6 | 3, 7 | 6 | 3, 7 | 6 | 3, 7 | |||
7 | 4, 7 | 7 | 3, 7 | 7 | 3, 7 | |||
8 | 4, 6 | 8 | 3, 7 | 8 | 3, 7 | |||
9 | 4, 6 | 9 | 3, 7 | 9 | 3, 7 | |||
10 | 4, 6 | 10 | 3, 7 | 10 | 4, 6 | |||
80 | 1 | 2, 8 | 80 | 1 | 0, 10 | 80 | 1 | 1, 9 |
2 | 3, 7 | 2 | 1, 9 | 2 | 2, 8 | |||
3 | 3, 7 | 3 | 2, 8 | 3 | 3, 7 | |||
4 | 3, 7 | 4 | 2, 8 | 4 | 3, 7 | |||
5 | 3, 7 | 5 | 3, 7 | 5 | 3, 7 | |||
6 | 4, 7 | 6 | 3, 7 | 6 | 3, 7 | |||
7 | 4, 6 | 7 | 3, 7 | 7 | 3, 7 | |||
8 | 4, 6 | 8 | 3, 7 | 8 | 4, 6 | |||
9 | 4, 6 | 9 | 3, 7 | 9 | 4, 6 | |||
10 | 4, 6 | 10 | 3, 7 | 10 | 4, 6 | |||
100 | 1 | 2, 8 | 100 | 1 | 1, 9 | 100 | 1 | 1, 9 |
2 | 3, 7 | 2 | 2, 8 | 2 | 2, 8 | |||
3 | 3, 7 | 3 | 2, 8 | 3 | 3, 7 | |||
4 | 3,7 | 4 | 3, 7 | 4 | 3, 7 | |||
5 | 4, 6 | 5 | 3, 7 | 5 | 3, 7 | |||
6 | 4, 6 | 6 | 3, 7 | 6 | 4, 6 | |||
7 | 4, 6 | 7 | 3, 7 | 7 | 4, 6 | |||
8 | 4, 6 | 8 | 3, 7 | 8 | 4, 6 | |||
9 | 4, 6 | 9 | 4, 7 | 9 | 4, 6 | |||
10 | 4, 6 | 10 | 4, 6 | 10 | 4, 6 |
Table 5.
Measured TpTe change (95% CI) | Measured QT change (95% CI) | Measured (QT)a (95%CI) | |||||||
---|---|---|---|---|---|---|---|---|---|
| |||||||||
n | 5 ms | 10 ms | 20 ms | 5 ms | 10 ms | 20 ms | 5 ms | 10 ms | 20 ms |
1 | -17, 27 | -12, 32 | -2, 42 | -27, 37 | -22, 42 | -12, 52 | -21, 31 | -16, 31 | -6, 46 |
2 | -10, 20 | -5, 25 | 5, 35 | -17, 27 | -12, 32 | -2, 42 | -13, 23 | -8, 23 | 2, 38 |
3 | -8, 18 | -3. 23 | 7, 33 | -13, 23 | -8, 28 | 2, 39 | -10, 20 | -5, 25 | 5, 35 |
4 | -6, 16 | -1, 21 | 9, 31 | -11, 21 | -6, 26 | 4, 36 | -8, 18 | -3, 23 | 7, 33 |
5 | -5, 15 | 0, 20 | 10, 30 | -9, 19 | -4, 24 | 6, 34 | -6, 16 | -1, 21 | 9, 31 |
10 | -2, 12 | 3, 17 | 13, 27 | -5, 15 | 0, 20 | 10, 30 | -3, 13 | 2, 18 | 12, 28 |
20 | 0, 10 | 5, 15 | 15, 25 | -2, 12 | 3, 17 | 13, 27 | -1, 11 | 4, 16 | 14, 26 |
Discussion
The TpTe and QT intervals from the 12-lead ECG have excellent short-term repeatability as the between-participant variability accounts for most of the total variability. Our mean values for TpTe were similar to those reported in a study of a younger population (92 ± 11 ms for women and 94±10 ms for men) by Haarmark et al (19). The results of our repeatability analysis were similar for TpTe and (TpTe)a (not shown). This similarity between TpTe and (TpTe)a was reported in a study of the electrocardiographic predictors of coronary heart disease and SCD in ARIC participants where investigators reported that TpTe in CVD-free participants was practically independent from heart rate (8).
Short-term variability in TpTe and QT like that of other ECG markers used to characterize ventricular repolarization is mostly explained by two main factors: biological variability and measurement error (6). Between-participant variability was the largest source of variability for TpTe and (QT)a. Within-visit variability accounted for slightly more variation than the between-visit for TpTe and (QT)a, which is likely related to the model assumption that the residual variation is due to within-visit variation and to the difficulty of partitioning within-visit from between-visit variation. This suggests that little variation was added when relating two ECGs from different visits compared to relating two ECGs from the same visit. The process-related variability was minimized by the standard ARIC protocol that was followed for electrode positioning, room condition, data collection and transfer (15).
The traditional measure of myocardial repolarization is the QT interval because its prognostic value is associated with arrhythmic risk, specifically torsades de pointes (TdP), a polymorphic ventricular tachycardia (20). TdP is the most serious adverse effect of drug therapy (21) and it is associated with rates of mortality between 10-17% (22). Because of these adverse associations, presence of QT prolongation has led to termination of development and withdraw from the market of many drugs, rejection of new drugs, and restrictions on prescribing new and old drugs (22, 23). Although QT prolongation has been widely used as a predictor of arrhythmic risk, it has more recently been described as being an imperfect measure (22). TpTe has been shown to be associated with SCD as the odds of prolonged TpTe were reported to be greater prior to and unrelated to SCD events, compared to controls from the same geographic area who had coronary artery disease (3). The limitations of QT prolongation as a surrogate of TdP and other VT and VF and as a measure to predict arrhythmic risk (22) are related to the interest in new measures of repolarization. TpTe has emerged as a novel measure of myocardial repolarization and a supplement to QT conveying information on arrhythmic risk (24).
To our knowledge, this is the first study to evaluate the short-term repeatability of TpTe from the standard 12-lead ECG. Another study using the same study population as ours, but without excluding participants with extreme outliers for TpTe and QT or evidence of an arrhythmia (n=63), reported that QT is a repeatable measure of ventricular repolarization given that between-participant variation (86%) was the main source of the variation between repeated measurements (5). The ICC for TpTe and QT were all close to unity and therefore more likely that the participants in our study and those in the study of QT repeatability displayed consistent measurements observed across visits compared to differences in measurements observed across participants.
The MDC estimate for TpTe was comparable to its SDs. Similarly, the MDC for (QT)a was comparable to its SD. Examination of the impact of repeated measurements on precision showed consistent results for TpTe and (QT)a, where gains in precision were seen with increasing numbers of enrolled participants.
The relatively small sample size is a limitation of our study. Another limitation was that our study design did not allow for an estimation of technician variation, however the variance due to technician is likely negligible as reported by Schroeder and Vaidean (6, 25). The strengths of our study include the standardized ECG measurement protocol conducted at two time points and the central training of the technicians.
The high repeatability and low MDC between measurements support the use of TpTe as a supplement to QT in observational and clinical studies to evaluate myocardial repolarization characteristics related to ventricular arrhythmia, heart failure and sudden cardiac death.
Acknowledgments
The Atherosclerosis Risk in Communities Study is carried out as a collaborative study supported by National Heart Lung, and Blood Institute contracts (HHSN268201100005C, HHSN268201100006C, HHSN268201100007C, HHSN268201100008C, HHSN268201100009C, HHSN268201100010C, HHSN268201100011C, and HHSN268201100012C). The authors thank the study participants for their important contribution. K.S.G. and M.L.S were supported by the National Heart Lung, and Blood Institute T32 training grant HL-007055.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content and all legal disclaimers that apply to the journal pertain.
References
- 1.Zhang Y, Post WS, Blasco-Colmenares E, et al. Electrocardiographic QT interval and mortality: a meta-analysis. Epidemiology. 2011;22(5):660–70. doi: 10.1097/EDE.0b013e318225768b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Kanters JK, Haarmark C, Vedel-Larsen E, et al. T(peak)T(end) interval in long QT syndrome. J Electrocardiol. 2008;41(6):603–8. doi: 10.1016/j.jelectrocard.2008.07.024. [DOI] [PubMed] [Google Scholar]
- 3.Panikkath R, Reinier K, Uy-Evanado A, et al. Prolonged Tpeak-to-tend interval on the resting ECG is associated with increased risk of sudden cardiac death. Circ Arrhythm Electrophysiol. 2011;4(4):441–7. doi: 10.1161/CIRCEP.110.960658. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Castro Hevia J, Antzelevitch C, Tornes Barzaga F, et al. Tpeak-Tend and Tpeak-Tend dispersion as risk factors for ventricular tachycardia/ventricular fibrillation in patients with the Brugada syndrome. J Am Coll Cardiol. 2006;47(9):1828–34. doi: 10.1016/j.jacc.2005.12.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Antzelevitch C. T peak-Tend interval as an index of transmural dispersion of repolarization. Eur J Clin Invest. 2001;31(7):555–7. doi: 10.1046/j.1365-2362.2001.00849.x. [DOI] [PubMed] [Google Scholar]
- 6.Vaidean GD, Schroeder EB, Whitsel EA, et al. Short-term repeatability of electrocardiographic spatial T-wave axis and QT interval. J Electrocardiol. 2005;38(2):139–47. doi: 10.1016/j.jelectrocard.2004.09.020. [DOI] [PubMed] [Google Scholar]
- 7.Roden DM. Drug-induced prolongation of the QT interval. N Engl J Med. 2004;350(10):1013–22. doi: 10.1056/NEJMra032426. [DOI] [PubMed] [Google Scholar]
- 8.Rautaharju PM, Zhang ZM, Warren J, et al. Electrocardiographic predictors of coronary heart disease and sudden cardiac deaths in men and women free from cardiovascular disease in the atherosclerosis risk in communities study. J Am Heart Assoc. 2013;2(3):e000061. doi: 10.1161/JAHA.113.000061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Rautaharju PM, Zhang ZM, Haisty WK, Jr, et al. Electrocardiographic Predictors of Incident Heart Failure in Men and Women Free From Manifest Cardiovascular Disease (from the Atherosclerosis Risk in Communities [ARIC] Study) Am J Cardiol. 2013 doi: 10.1016/j.amjcard.2013.05.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Yan GX, Antzelevitch C. Cellular basis for the normal T wave and the electrocardiographic manifestations of the long-QT syndrome. Circulation. 1998;98(18):1928–36. doi: 10.1161/01.cir.98.18.1928. [DOI] [PubMed] [Google Scholar]
- 11.Xia Y, Liang Y, Kongstad O, et al. In vivo validation of the coincidence of the peak and end of the T wave with full repolarization of the epicardium and endocardium in swine. Heart Rhythm. 2005;2(2):162–9. doi: 10.1016/j.hrthm.2004.11.011. [DOI] [PubMed] [Google Scholar]
- 12.Zhu TG, Patel C, Martin S, et al. Ventricular transmural repolarization sequence: its relationship with ventricular relaxation and role in ventricular diastolic function. Eur Heart J. 2009;30(3):372–80. doi: 10.1093/eurheartj/ehn585. [DOI] [PubMed] [Google Scholar]
- 13.Opthof T, Coronel R, Wilms-Schopman FJ, et al. Dispersion of repolarization in canine ventricle and the electrocardiographic T wave: Tp-e interval does not reflect transmural dispersion. Heart Rhythm. 2007;4(3):341–8. doi: 10.1016/j.hrthm.2006.11.022. [DOI] [PubMed] [Google Scholar]
- 14.Xia Y, Liang Y, Kongstad O, et al. Tpeak-Tend interval as an index of global dispersion of ventricular repolarization: evaluations using monophasic action potential mapping of the epi- and endocardium in swine. J Interv Card Electrophysiol. 2005;14(2):79–87. doi: 10.1007/s10840-005-4592-4. [DOI] [PubMed] [Google Scholar]
- 15.The Atherosclerosis Risk in Communities (ARIC) Study: design and objectives. The ARIC investigators. Am J Epidemiol. 1989;129(4):687–702. [PubMed] [Google Scholar]
- 16.Rautaharju PM, Wolf HK, Eifler WJ, et al. A simple procedure for positioning precordial ECG and VCG electrodes using an electrode locator. J Electrocardiol. 1976;9(1):35–40. doi: 10.1016/s0022-0736(76)80007-6. [DOI] [PubMed] [Google Scholar]
- 17.Hnatkova K, Gang Y, Batchvarov VN, et al. Precision of QT interval measurement by advanced electrocardiographic equipment. Pacing Clin Electrophysiol. 2006;29(11):1277–84. doi: 10.1111/j.1540-8159.2006.00532.x. [DOI] [PubMed] [Google Scholar]
- 18.Bazett J. An analysis of time relations of electrocardiograms. Heart. 1920;7:353–367. Heart. 1920;7:353-67. [Google Scholar]
- 19.Haarmark C, Graff C, Andersen MP, et al. Reference values of electrocardiogram repolarization variables in a healthy population. J Electrocardiol. 2010;43(1):31–9. doi: 10.1016/j.jelectrocard.2009.08.001. doi: http://dx.doi.org.libproxy.lib.unc.edu/10.1016/j.jelectrocard.2009.08.001. [DOI] [PubMed] [Google Scholar]
- 20.Khan IA, Win MT, Bali AJ, et al. Torsades de pointes: a case with multiple variables. Am J Emerg Med. 1999;17(1):80–5. doi: 10.1016/s0735-6757(99)90025-6. [DOI] [PubMed] [Google Scholar]
- 21.Morganroth J. Relations of QTc prolongation on the electrocardiogram to torsades de pointes: definitions and mechanisms. Am J Cardiol. 1993;72(6):10B–3B. doi: 10.1016/0002-9149(93)90033-9. [DOI] [PubMed] [Google Scholar]
- 22.Shah RR, Hondeghem LM. Refining detection of drug-induced proarrhythmia: QT interval and TRIaD. Heart Rhythm. 2005;2(7):758–72. doi: 10.1016/j.hrthm.2005.03.023. [DOI] [PubMed] [Google Scholar]
- 23.Food and Drug Administration, HHS. International Conference on Harmonisation; guidance on E14 Clinical Evaluation of QT/QTc Interval Prolongation and Proarrhythmic Potential for Non-Antiarrhythmic Drugs; availability. Notice. Fed Regist. 2005;70(202):61134–5. [PubMed] [Google Scholar]
- 24.Huikuri HV, Castellanos A, Myerburg RJ. Sudden death due to cardiac arrhythmias. N Engl J Med. 2001;345(20):1473–82. doi: 10.1056/NEJMra000650. [DOI] [PubMed] [Google Scholar]
- 25.Schroeder EB, Whitsel EA, Evans GW, et al. Repeatability of heart rate variability measures. J Electrocardiol. 2004;37(3):163–72. doi: 10.1016/j.jelectrocard.2004.04.004. [DOI] [PubMed] [Google Scholar]