TABLE III.

Time (min) for Sensors to Reach 90% Steady State Current in Response to Step Changes in Glucose Concentration

Time (min) to Reach 90% Steady State Current
0 → 100a 100 → 200a 200 → 400a 400 → 0a
Bare sensor (n = 8) 2 ± 0 2 ± 0 2 ± 0 15 ± 1
Small pore coating (n = 4) 3 ± 1 3 ± 0 4 ± 0 13 ± 3
Large pore coating (n = 4) 13 ± 3 20 ± 1 26 ± 1 >45b

Data presented as mean ± SE.

a

Glucose challenge (mg/dL).

b

Time exceeded the experimental time interval.