Table 1.

Patient-based analysis: intraobserver and interobserver variability in the sum longest diameter (SLD) of target lesions for baseline measurements and for the percent change at the first follow-up scan.

Intraobserver variability of SLD (Two measurements by Radiologist 1)
CCC (95%CI) Mean relative difference(%) 95% limits of agreement(%)
Baseline (n=71) 0.9984 (0.9974–0.9990) −0.365 −7.90, 7.17
%change on follow-up (n=69) 0.9747 (0.9597–0.9841) 0.282 −7.30, 7.86
Interobserver variability of SLD (Measurements by Radiologist 1 vs. Radiologist 2)
CCC (95%CI) Mean relative difference(%) 95% limits of agreement(%)
Baseline (n=71) 0.9875 (0.9800–0.9921) −0.173 −17.0, 16.6
%change on follow-up (n=69) 0.8969 (0.8389–0.9348) −0.444 −16.3, 15.4