Table 3.
GE-QC program water test tolerances.
Scanner model | Mean CT# water (HU) | Std dev water (HU) | Unif max diff (HU) |
---|---|---|---|
LS+ | ±3 | 2.6–3.4 | ±3 |
DST | ±3 | 2.6–3.4 | ±3 |
LS-16/DRX/DSTE | ±3 | 2.9–3.5 | ±3 |
VCT/DVCT | ±3 | 3.8–4.8 | ±3 |
GE-QC program water test tolerances.
Scanner model | Mean CT# water (HU) | Std dev water (HU) | Unif max diff (HU) |
---|---|---|---|
LS+ | ±3 | 2.6–3.4 | ±3 |
DST | ±3 | 2.6–3.4 | ±3 |
LS-16/DRX/DSTE | ±3 | 2.9–3.5 | ±3 |
VCT/DVCT | ±3 | 3.8–4.8 | ±3 |