S-QC program water test tolerances.

Scanner model Mean CT# water (HU) Std dev water (HU) Unif max diff (HU)
Sensation open ±4.0 9.72–11.88 ±4.0
Sensation 16 ±4.0 6.35–7.76 ±4.0
Sensation 64 ±4.0 10.89–13.31 ±4.0
Emotion 6 ±4.0 10.8–13.2 ±4.0
Emotion 16 ±4.0 9.27–7.76 ±4.0
1