Skip to main content
. 2014 Apr 4;51(6):419–424. doi: 10.1136/jmedgenet-2014-102360

Table 1.

The results of 14 laboratories on 78 samples

  Laboratories
A B C D E F G H I J K L M N
RP-PCR results
 GGGGCC-repeat expansion
  Yes 46 46 46 46 42 48 45 43 43 45 34 44 41 43
  No 32 32 32 32 32 28 33 32 32 30 33 32 33 28
  Q 0 0 0 0 4 2 0 3 3 3 11 2 4 7
 False positive 0 0 0 0 0 2 0 0 0 2 1 0 0 0
 False negative 0 0 0 0 1 0 1 0 0 0 2 0 1 0
 Sensitivity (%) 100 100 100 100 91.3 100 97.8 93.5 93.5 93.5 71.7 95.7 89.1 93.5
 Specificity (%) 100 100 100 100 96.8 87.5 100 100 100 93.8 96.9 100 100 87.5
 Unclassified (%) 0 0 0 0 5.1 2.6 0 3.8 3.8 3.8 14.1 2.6 5.1 9.0
Amplicon-length analysis results
 Numbers of amplicon
  0 0 0 1 0 0 0 0 0 0 0 0 N.D. N.D. N.D.
  1 54 54 55 56 56 54 54 51 56 54 51 N.D. N.D. N.D.
  2 24 24 22 22 22 24 24 24 22 24 24 N.D. N.D. N.D.
  Q 0 0 0 0 0 0 0 3 0 0 3 N.D. N.D. N.D.
RP-PCR+amplicon-length analysis results
 GGGGCC-repeat expansion
  Yes 46 46 46 46 46 48 45 43 43 43 34 44 41 43
  No 32 32 32 32 32 29 33 32 32 30 34 32 33 28
  Q 0 0 0 0 0 1 0 3 3 5 10 2 4 7
 False positive 0 0 0 0 0 2 0 0 0 0 0 0 0 0
 False negative 0 0 0 0 0 0 1 0 0 0 2 0 1 0
 Sensitivity (%) 100 100 100 100 100 100 97.8 93.5 93.5 93.5 73.9 95.7 89.1 93.5
 Specificity (%) 100 100 100 100 100 90.6 100 100 100 93.8 100 100 100 87.5
 Unclassified (%) 0 0 0 0 0 1.3 0 3.8 3.8 6.4 12.8 2.6 5.1 9.0

No, sample numbers without large GGGGCC-repeat expansion; N.D., not determined; Q, sample numbers of undecided or unclassified samples Yes: sample numbers with large GGGGCC-repeat expansion.