Skip to main content
. Author manuscript; available in PMC: 2014 May 26.
Published in final edited form as: Clin Chim Acta. 2013 Mar 13;421:91–97. doi: 10.1016/j.cca.2013.03.003

Table 1.

Intra-day and inter-day accuracies and precisions as well as extraction recoveries for S-adenosylmethionine and S-adenosylhomocysteine.

Analyte Quality control concentration Intra-day accuracy ± coefficient of variation Inter-day accuracy ± coefficient of variation Recovery ± coefficient of variation




nmol/L [%] [%] [%]
SAM 37.5 93.4 ± 6.6 96.5 ± 9.2 99.7 ± 7.2
75 96.6 ± 5.3 99.3 ± 8.6 99.8 ± 2.4
150 91.9 ± 4.4 97.4 ± 8.9 99.2 ± 2.1
300 105.5 ± 2.4 101.5 ± 8.7 104.6 ± 3.0
900 98.9 ± 7.1 103.9 ± 8.1 103.3 ± 2.2
SAH 37.5 95.0 ± 8.1 98.9 ± 8.7 96.7 ± 5.8
75 90.3 ± 4.3 99.3 ± 9.6 95.4 ± 2.3
150 89.7 ± 3.0 97.9 ± 9.8 92.7 ± 3.4
300 98.5 ± 5.3 99.0 ± 9.7 97.1 ± 1.1
900 90.3 ± 2.5 98.9 ± 8.4 103.5 ± 4.4