Published in final edited form as: Nat Protoc. 2012 Jun 7;7(7):1247–1259. doi: 10.1038/nprot.2012.051

Table 2.

Troubleshooting table.

Step Problem Possible reason Solution
26 Defect in bonding Mishandling in bonding process Press the defective region using a pair of tweezers until the unbounded PDMS surface adheres to the glass surface.
Weak bonding Weak plasma treatment Increase plasma treatment time or place the bonded device at 80 °C in an oven for several hours.
41 Collagen solution flows out into the media channel Hydrophobicity of the device not yet recovered to the required level Let device dry for a longer time (e.g. 48 hrs after PDL coating) in an 80 °C oven.
54 Non-confluent endothelial monolayer on one day Low hMVECs seeding density Wait one more day and increase cell seeding density.
Unhealthy hMVECs Poor surface coating Use fresh surface coating solution.
Media evaporation in reservoirs Maintain a stable humidity in the incubator and add more medium into the reservoirs.