Table 7.
ET, EU/l | Bacteria, CFU/ml | |
---|---|---|
Facility A | ||
Tap water | 8,076 | 200 |
RO water | 57 | 100 |
Dialysate 1 | 741 | 500 |
Dialysate 2 | 903 | 800 |
Facility B | ||
Tap water | 378 | 70 |
RO water | 2.9 | 300 |
Dialysate 1 | 38 | 150 |
Dialysate 2 | 49 | 100 |
Facility C | ||
Tap water | 1,468 | 110 |
RO water | 94 | 22 |
Dialysate 1 | 201 | 600 |
Dialysate 2 | 395 | 1,000 |
Dialysate 3a | 2,295 | 1,000 |
Facility D | ||
Tap water | 1,663 | 110 |
RO water 1 | 7.3 | 120 |
RO water 2 | 7.8 | 100 |
Dialysate 1 | 309 | 800 |
Dialysate 2 | 246 | 800 |
At a console with an ET-retentive filter.