Figure 3.

Figure 3

Titration series of PSA in buffer (PBS) solution at three different concentrations of the capturing antibody H117 (15 µgmL−1, µgmL−1 and 154 µgmL−1). The LOD was found to be 2.5 ngmL−1 when the capturing antibody was 15 µgmL−1 and was reduced to 80 pgmL−1 and 800 fgmL−1 when the capturing antibody concentrations were 35 µgmL−1 and 154 µgmL−1, respectively. Signal of negative control (N) increases significantly at the higher concentration of H117 (154µgmL−1).