Published in final edited form as: Magn Reson Med. 2013 Jul 15;71(6):2197–2205. doi: 10.1002/mrm.24863

Table 1.

Scanning Parameters and Post-Processing Setup for the Different Field Strengths and Experiments

B0 9.4 T 7.0 T 7.0 T 3.0 T
Scan Ex Vivo In Vivo Ex Vivo Ex Vivo
TR/TE (ms) 40/14 27/15 26/13 45/28
BW (Hz/pixel) ~43 90 80 40
Flip angle (deg) 15 15 15 15
Resolution (μm) ~100 250 210 375
Matrix, slices 2563–3843 384 × 384, 144 384 × 384, 120 320 × 240, 88
Averages 1 3 2 1
Accelerations GRAPPA = 2, Partial Fourier = 6/8a GRAPPA = 2, Partial Fourier = 6/8a
Flow compensation Yes Yes Yes
Scan time (min) 43–98 22:41 11:07 15:54
Post-processing In-house, according to Haacke (4) Vendor-provided and in-house Vendor-provided Vendor-provided
a

Both slice and phase partial Fourier.