Table 2.

Ratio of Maximum Local to Whole-Head Average SAR for a Human Head Model in a Volume Coil at Several Frequencies Compared to the Ratios in the Regulatory Limits

Frequency SAR1g/SARW SAR10g/SARW
64 MHz 6.1 3.4
128 MHz 5.2 3.2
200 MHz 5.6 3.5
300 MHz 5.4 4.1
340 MHz 6.7 5.0
400 MHz 6.0 4.4
FDA limits 2.7
IEC limits 3.12

FDA = Food and Drug Administration, IEC = International Electrotechnical Commission.