Table 3.

Ratio of Maximum Local to Whole-Body Average SAR for Two Different Subjects in a Whole-Body Volume Coil Compared to the Ratios in Corresponding Regulatory Limits*

Subject SAR1g/SARW SAR10g/SARW
64 MHz 1 18 10
2 16 11
128 MHz 1 18 10
2 16 13
FDA limits 2
IEC limits 5
*

Results derived recently from calculations for a previous publication (8).