Table 3.
Ratio of Maximum Local to Whole-Body Average SAR for Two Different Subjects in a Whole-Body Volume Coil Compared to the Ratios in Corresponding Regulatory Limits*
Subject | SAR1g/SARW | SAR10g/SARW | |
---|---|---|---|
64 MHz | 1 | 18 | 10 |
2 | 16 | 11 | |
128 MHz | 1 | 18 | 10 |
2 | 16 | 13 | |
FDA limits | 2 | – | |
IEC limits | – | 5 |
Results derived recently from calculations for a previous publication (8).
FDA = Food and Drug Administration, IEC = International Electrotechnical Commission.