. 2014 Jun 3;9(6):e98487. doi: 10.1371/journal.pone.0098487

Table 2. Improvements to Innocor gas analyser rise time.

Condition Mean T90 (ms) SD T90 (ms) N repeats Mean (SD) FGD (ms)
1 Standard system (170 cm gas sample line) 154.6 0.95 41 1376 (3.7)
2 Oxigraf removed 125.5 1.2 19 1026 (2.3)
3 Nafion gas sample line length 150 cm 121.1 1.96 25 932 (2.6)
130 cm 119.0 2.12 23 860 (3.7)
105 cm 115.9 2.36 44 782 (3.3)
90 cm 113.6 2.11 20 708 (2.2)
4 Custom-optimised connectors 88.6 1.6 28 671 (3.6)
5 Photoacoustic gas analyser only 55.9 0.65 8 -

Effect of modifications to Innocor hardware on rise time (T90) and flow gas delay of SF6 signal. Steps were performed sequentially in the numbered order. FGD - flow gas delay.