. 2014 Jun 3;9(6):e98487. doi: 10.1371/journal.pone.0098487

Table 3. Accuracy of standard and enhanced Innocor gas analyser.

System T90 (ms) FRC (L) Mean (SD) percentage error in FRC
Slow ventilation (10–12 min−1) Medium (20–22 min−1) Fast (31–32 min−1)
Fast 88 0.521 1.87 (0.77) 1.48 (0.73) 0.79 (0.27)
1.011 1.15 (1.2) 0.94 (0.46) 0.40 (0.18)
2.001 −0.06 (0.63) 0.04 (0.15) 0.21 (0.95)
Slow 154 0.521 0.07 (0.30) −2.72 (0.81) −8.33 (0.36)
1.011 −0.30 (0.87) −4.58 (0.24) −8.89 (0.21)
2.001 −1.65 (0.87) −2.67 (0.12) −5.91 (0.34)

Summary of error in functional residual capacity (FRC) calculated by multiple breath washout for a lung model at different FRC and ventilation rates. FRC are shown excluding the equipment dead space of 59 ml, and error is quoted as a percentage of the lung model FRC only (excluding deadspace). Precise ventilation rates varied with the lung model settings specific to each FRC and were 10.9, 21.8, 31.6 min−1 for 0.5 L FRC, 10.8, 21.7, 31.5 min−1 for 1 L FRC, and 11.8, 20.5, 31.3 min−1 for 2 L FRC. T90: 10–90% rise time of the gas analyser.