Table 3. Accuracy of standard and enhanced Innocor gas analyser.
System T90 (ms) | FRC (L) | Mean (SD) percentage error in FRC | ||
Slow ventilation (10–12 min−1) | Medium (20–22 min−1) | Fast (31–32 min−1) | ||
Fast 88 | 0.521 | 1.87 (0.77) | 1.48 (0.73) | 0.79 (0.27) |
1.011 | 1.15 (1.2) | 0.94 (0.46) | 0.40 (0.18) | |
2.001 | −0.06 (0.63) | 0.04 (0.15) | 0.21 (0.95) | |
Slow 154 | 0.521 | 0.07 (0.30) | −2.72 (0.81) | −8.33 (0.36) |
1.011 | −0.30 (0.87) | −4.58 (0.24) | −8.89 (0.21) | |
2.001 | −1.65 (0.87) | −2.67 (0.12) | −5.91 (0.34) |
Summary of error in functional residual capacity (FRC) calculated by multiple breath washout for a lung model at different FRC and ventilation rates. FRC are shown excluding the equipment dead space of 59 ml, and error is quoted as a percentage of the lung model FRC only (excluding deadspace). Precise ventilation rates varied with the lung model settings specific to each FRC and were 10.9, 21.8, 31.6 min−1 for 0.5 L FRC, 10.8, 21.7, 31.5 min−1 for 1 L FRC, and 11.8, 20.5, 31.3 min−1 for 2 L FRC. T90: 10–90% rise time of the gas analyser.